A cadaveric precision and accuracy analysis of augmented reality–mediated percutaneous pedicle implant insertion

J Neurosurg Spine 34:316–324, 2021

Augmented reality–mediated spine surgery (ARMSS) is a minimally invasive novel technology that has the potential to increase the efficiency, accuracy, and safety of conventional percutaneous pedicle screw insertion methods. Visual 3D spinal anatomical and 2D navigation images are directly projected onto the operator’s retina and superimposed over the surgical field, eliminating field of vision and attention shift to a remote display. The objective of this cadaveric study was to assess the accuracy and precision of percutaneous ARMSS pedicle implant insertion.

METHODS Instrumentation was placed in 5 cadaveric torsos via ARMSS with the xvision augmented reality headmounted display (AR-HMD) platform at levels ranging from T5 to S1 for a total of 113 total implants (93 pedicle screws and 20 Jamshidi needles). Postprocedural CT scans were graded by two independent neuroradiologists using the Gertzbein- Robbins scale (grades A–E) for clinical accuracy. Technical precision was calculated using superimposition analysis employing the Medical Image Interaction Toolkit to yield angular trajectory (°) and linear screw tip (mm) deviation from the virtual pedicle screw position compared with the actual pedicle screw position on postprocedural CT imaging.

RESULTS The overall implant insertion clinical accuracy achieved was 99.1%. Lumbosacral and thoracic clinical accuracies were 100% and 98.2%, respectively. Specifically, among all implants inserted, 112 were noted to be Gertzbein- Robbins grade A or B (99.12%), with only 1 medial Gertzbein-Robbins grade C breach (> 2-mm pedicle breach) in a thoracic pedicle at T9. Precision analysis of the inserted pedicle screws yielded a mean screw tip linear deviation of 1.98 mm (99% CI 1.74–2.22 mm) and a mean angular error of 1.29° (99% CI 1.11°–1.46°) from the projected trajectory. These data compare favorably with data from existing navigation platforms and regulatory precision requirements mandating that linear and angular deviation be less than 3 mm (p < 0.01) and 3° (p < 0.01), respectively.

CONCLUSIONS Percutaneous ARMSS pedicle implant insertion is a technically feasible, accurate, and highly precise method.

 

A wearable mixed-reality holographic computer for guiding external ventricular drain insertion at the bedside

J Neurosurg 131:1599–1606, 2019

The goal of this study was to explore the feasibility and accuracy of using a wearable mixed-reality holographic computer to guide external ventricular drain (EVD) insertion and thus improve on the accuracy of the classic freehand insertion method for EVD insertion. The authors also sought to provide a clinically applicable workflow demonstration.

METHODS Pre- and postoperative CT scanning were performed routinely by the authors for every patient who needed EVD insertion. Hologram-guided EVD placement was prospectively applied in 15 patients between August and November 2017. During surgical planning, model reconstruction and trajectory calculation for each patient were completed using preoperative CT. By wearing a Microsoft HoloLens, the neurosurgeon was able to visualize the preoperative CT-generated holograms of the surgical plan and perform EVD placement by keeping the catheter aligned with the holographic trajectory. Fifteen patients who had undergone classic freehand EVD insertion were retrospectively included as controls. The feasibility and accuracy of the hologram-guided technique were evaluated by comparing the time required, number of passes, and target deviation for hologram-guided EVD placement with those for classic freehand EVD insertion.

RESULTS Surgical planning and hologram visualization were performed in all 15 cases in which EVD insertion involved holographic guidance. No adverse events related to the hologram-guided procedures were observed. The mean ± SD additional time before the surgical part of the procedure began was 40.20 ± 10.74 minutes. The average number of passes was 1.07 ± 0.258 in the holographic guidance group, compared with 2.33 ± 0.98 in the control group (p < 0.01). The mean target deviation was 4.34 ± 1.63 mm in the holographic guidance group and 11.26 ± 4.83 mm in the control group (p < 0.01).

CONCLUSIONS This study demonstrates the use of a head-mounted mixed-reality holographic computer to successfully perform hologram-assisted bedside EVD insertion. A full set of clinically applicable workflow images is presented to show how medical imaging data can be used by the neurosurgeon to visualize patient-specific holograms that can intuitively guide hands-on operation. The authors also provide preliminary confirmation of the feasibility and accuracy of this hologram-guided EVD insertion technique.

Augmented reality–assisted pedicle screw insertion

J Neurosurg Spine 31:139–146, 2019

Augmented reality (AR) is a novel technology that has the potential to increase the technical feasibility, accuracy, and safety of conventional manual and robotic computer-navigated pedicle insertion methods. Visual data are directly projected to the operator’s retina and overlaid onto the surgical field, thereby removing the requirement to shift attention to a remote display. The objective of this study was to assess the comparative accuracy of AR-assisted pedicle screw insertion in comparison to conventional pedicle screw insertion methods.

METHODS Five cadaveric male torsos were instrumented bilaterally from T6 to L5 for a total of 120 inserted pedicle screws. Postprocedural CT scans were obtained, and screw insertion accuracy was graded by 2 independent neuroradiologists using both the Gertzbein scale (GS) and a combination of that scale and the Heary classification, referred to in this paper as the Heary-Gertzbein scale (HGS). Non-inferiority analysis was performed, comparing the accuracy to freehand, manual computer-navigated, and robotics-assisted computer-navigated insertion accuracy rates reported in the literature. User experience analysis was conducted via a user experience questionnaire filled out by operators after the procedures.

RESULTS The overall screw placement accuracy achieved with the AR system was 96.7% based on the HGS and 94.6% based on the GS. Insertion accuracy was non-inferior to accuracy reported for manual computer-navigated pedicle insertion based on both the GS and the HGS scores. When compared to accuracy reported for robotics-assisted computer-navigated insertion, accuracy achieved with the AR system was found to be non-inferior when assessed with the GS, but superior when assessed with the HGS. Last, accuracy results achieved with the AR system were found to be superior to results obtained with freehand insertion based on both the HGS and the GS scores. Accuracy results were not found to be inferior in any comparison. User experience analysis yielded “excellent” usability classification.

CONCLUSIONS AR-assisted pedicle screw insertion is a technically feasible and accurate insertion method.

 

App-assisted external ventricular drain insertion

app-assisted-external-ventricular-drain-insertion

J Neurosurg 125:754–758, 2016

The freehand technique for insertion of an external ventricular drain (EVD) is based on fixed anatomical landmarks and does not take individual variations into consideration. A patient-tailored approach based on augmented-reality techniques using devices such as smartphones can address this shortcoming. The Sina neurosurgical assist (Sina) is an Android mobile device application (app) that was designed and developed to be used as a simple intraoperative neurosurgical planning aid. It overlaps the patient’s images from previously performed CT or MRI studies on the image seen through the device camera.

The device is held by an assistant who aligns the images and provides information about the relative position of the target and EVD to the surgeon who is performing EVD insertion. This app can be used to provide guidance and continuous monitoring during EVD placement.

The author describes the technique of Sina-assisted EVD insertion into the frontal horn of the lateral ventricle and reports on its clinical application in 5 cases as well as the results of ex vivo studies of ease of use and precision. The technique has potential for further development and use with other augmented-reality devices.

Augmented reality–guided neurosurgery: accuracy and intraoperative application of an image projection technique

Augmented reality–guided neurosurgery- accuracy and intraoperative application of an image projection technique

J Neurosurg 123:206–211, 2015

An augmented reality system has been developed for image-guided neurosurgery to project images with regions of interest onto the patient’s head, skull, or brain surface in real time. The aim of this study was to evaluate system accuracy and to perform the first intraoperative application.

Methods Images of segmented brain tumors in different localizations and sizes were created in 10 cases and were projected to a head phantom using a video projector. Registration was performed using 5 fiducial markers. After each registration, the distance of the 5 fiducial markers from the visualized tumor borders was measured on the virtual image and on the phantom. The difference was considered a projection error. Moreover, the image projection technique was intraoperatively applied in 5 patients and was compared with a standard navigation system.

Results Augmented reality visualization of the tumors succeeded in all cases. The mean time for registration was 3.8 minutes (range 2–7 minutes). The mean projection error was 0.8 ± 0.25 mm. There were no significant differences in accuracy according to the localization and size of the tumor. Clinical feasibility and reliability of the augmented reality system could be proved intraoperatively in 5 patients (projection error 1.2 ± 0.54 mm).

Conclusions The augmented reality system is accurate and reliable for the intraoperative projection of images to the head, skull, and brain surface. The ergonomic advantage of this technique improves the planning of neurosurgical procedures and enables the surgeon to use direct visualization for image-guided neurosurgery.

The use of simulation in neurosurgical education and training

Simulation in neurosurgical training

J Neurosurg 121:228–246, 2014

There is increasing evidence that simulation provides high-quality, time-effective training in an era of resident duty-hour restrictions. Simulation may also permit trainees to acquire key skills in a safe environment, important in a specialty such as neurosurgery, where technical error can result in devastating consequences. The authors systematically reviewed the application of simulation within neurosurgical training and explored the state of the art in simulation within this specialty. To their knowledge this is the first systematic review published on this topic to date.

Methods. The authors searched the Ovid MEDLINE, Embase, and PsycINFO databases and identified 4101 articles; 195 abstracts were screened by 2 authors for inclusion. The authors reviewed data on study population, study design and setting, outcome measures, key findings, and limitations.

Results. Twenty-eight articles formed the basis of this systematic review. Several different simulators are at the neurosurgeon’s disposal, including those for ventriculostomy, neuroendoscopic procedures, and spinal surgery, with evidence for improved performance in a range of procedures. Feedback from participants has generally been favorable. However, study quality was found to be poor overall, with many studies hampered by nonrandomized design, presenting normal rather than abnormal anatomy, lack of control groups and long-term follow-up, poor study reporting, lack of evidence of improved simulator performance translating into clinical benefit, and poor reliability and validity evidence. The mean Medical Education Research Study Quality Instrument score of included studies was 9.21 ± 1.95 (± SD) out of a possible score of 18.

Conclusions. The authors demonstrate qualitative and quantitative benefits of a range of neurosurgical simulators but find significant shortfalls in methodology and design. Future studies should seek to improve study design and reporting, and provide long-term follow-up data on simulated and ideally patient outcomes.

Novel augmented reality–assisted percutaneous vertebroplasty

A novel 3D guidance system using augmented reality for percutaneous vertebroplasty

J Neurosurg Spine 19:492–501, 2013

Augmented reality (AR) is an imaging technology by which virtual objects are overlaid onto images of real objects captured in real time by a tracking camera. This study aimed to introduce a novel AR guidance system called virtual protractor with augmented reality (VIPAR) to visualize a needle trajectory in 3D space during percutaneous vertebroplasty (PVP). The AR system used for this study comprised a head-mount display (HMD) with a tracking camera and a marker sheet. An augmented scene was created by overlaying the preoperatively generated needle trajectory path onto a marker detected on the patient using AR software, thereby providing the surgeon with augmented views in real time through the HMD. The accuracy of the system was evaluated by using a computer-generated simulation model in a spine phantom and also evaluated clinically in 5 patients.

In the 40 spine phantom trials, the error of the insertion angle (EIA), defined as the difference between the attempted angle and the insertion angle, was evaluated using 3D CT scanning. Computed tomography analysis of the 40 spine phantom trials showed that the EIA in the axial plane significantly improved when VIPAR was used compared with when it was not used (0.96° ± 0.61° vs 4.34° ± 2.36°, respectively). The same held true for EIA in the sagittal plane (0.61° ± 0.70° vs 2.55° ± 1.93°, respectively). In the clinical evaluation of the AR system, 5 patients with osteoporotic vertebral fractures underwent VIPARguided PVP from October 2011 to May 2012. The postoperative EIA was evaluated using CT.

The clinical results of the 5 patients showed that the EIA in all 10 needle insertions was 2.09° ± 1.3° in the axial plane and 1.98° ± 1.8° in the sagittal plane. There was no pedicle breach or leakage of polymethylmethacrylate.

VIPAR was successfully used to assist in needle insertion during PVP by providing the surgeon with an ideal insertion point and needle trajectory through the HMD. The findings indicate that AR guidance technology can become a useful assistive device during spine surgeries requiring percutaneous procedures.