Use of the Airo mobile intraoperative CT system versus the O-arm for transpedicular screw fixation in the thoracic and lumbar spine

J Neurosurg Spine 29:397–406, 2018

Navigation-enabling technology such as 3D-platform (O-arm) or intraoperative mobile CT (iCT-Airo) systems for use in spinal surgery has considerably improved accuracy over that of traditional fluoroscopy-guided techniques during pedicular screw positioning. In this study, the authors compared 2 intraoperative imaging systems with navigation, available in their neurosurgical unit, in terms of the accuracy they provided for transpedicular screw fixation in the thoracic and lumbar spine.

METHODS The authors performed a retrospective analysis of clinical and surgical data of 263 consecutive patients who underwent thoracic and lumbar spine screw placement in the same center. Data on 97 patients who underwent surgery with iCT-Airo navigation (iCT-Airo group) and 166 with O-arm navigation (O-arm group) were analyzed. Most patients underwent surgery for a degenerative or traumatic condition that involved thoracic and lumbar pedicle screw fixation using an open or percutaneous technique. The primary endpoint was the proportion of patients with at least 1 screw not correctly positioned according to the last intraoperative image. Secondary endpoints were the proportion of screws that were repositioned during surgery, the proportion of patients with a postoperative complication related to screw malposition, surgical time, and radiation exposure. A blinded radiologist graded screw positions in the last intraoperative image according to the Heary classification (grade 1–3 screws were considered correctly placed).

RESULTS A total of 1361 screws placed in 97 patients in the iCT-Airo group (503 screws) and in 166 in the O-arm group (858 screws) were graded. Of those screws, 3 (0.6%) in the iCT-Airo group and 4 (0.5%) in the O-arm group were misplaced. No statistically significant difference in final accuracy between these 2 groups or in the subpopulation of patients who underwent percutaneous surgery was found. Three patients in the iCT-Airo group (3.1%, 95% CI 0%–6.9%) and 3 in the O-arm group (1.8%, 95% CI 0%–4.0%) had a misplaced screw (Heary grade 4 or 5). Seven (1.4%) screws in the iCT-Airo group and 37 (4.3%) in the O-arm group were repositioned intraoperatively (p = 0.003). One patient in the iCT-Airo group and 2 in the O-arm group experienced postoperative neurological deficits related to hardware malposition. The mean surgical times in both groups were similar (276 [iCT-Airo] and 279 [O-arm] minutes). The mean exposure to radiation in the iCT-Airo group was significantly lower than that in the O-arm group (15.82 vs 19.12 mSv, respectively; p = 0.02).

CONCLUSIONS Introduction of a mobile CT scanner reduced the rate of screw repositioning, which enhanced patient safety and diminished radiation exposure for patients, but it did not improve overall accuracy compared to that of a mobile 3D platform.

 

Interactive presurgical simulation applying 3D techniques

Presurgical simulation 3Dimaging

J Neurosurg 119:94–105, 2013

In this paper, the authors’ goal was to report their novel presurgical simulation method applying interactive virtual simulation (IVS) using 3D computer graphics (CG) data and microscopic observation of color-printed plaster models based on these CG data in surgery for skull base and deep tumors.

Methods. For 25 operations in 23 patients with skull base or deep intracranial tumors (meningiomas, schwannomas, epidermoid tumors, chordomas, and others), the authors carried out presurgical simulation based on 3D CG data created by image analysis for radiological data. Interactive virtual simulation was performed by modifying the 3D CG data to imitate various surgical procedures, such as bone drilling, brain retraction, and tumor removal, with manipulation of a haptic device. The authors also produced color-printed plaster models of modified 3D CG data by a selective laser sintering method and observed them under the operative microscope.

Results. In all patients, IVS provided detailed and realistic surgical perspectives of sufficient quality, thereby allowing surgeons to determine an appropriate and feasible surgical approach. Surgeons agreed that in 44% of the 25 operations IVS showed high utility (as indicated by a rating of “prominent”) in comprehending 3D microsurgical anatomies for which reconstruction using only 2D images was complicated. Microscopic observation of color-printed plaster models in 12 patients provided further utility in confirming realistic surgical anatomies.

Conclusions. The authors’ presurgical simulation method applying advanced 3D imaging and modeling techniques provided a realistic environment for practicing microsurgical procedures virtually and enabled the authors to ascertain complex microsurgical anatomy, to determine the optimal surgical strategies, and also to efficiently educate neurosurgical trainees, especially during surgery for skull base and deep tumors.

Interactive presurgical simulation applying 3D techniques

3D CG data and the color-printed plaster model for a left tentorial meningioma

J Neurosurg 119:94–105, 2013

In this paper, the authors’ goal was to report their novel presurgical simulation method applying interactive virtual simulation (IVS) using 3D computer graphics (CG) data and microscopic observation of color-printed plaster models based on these CG data in surgery for skull base and deep tumors.

Methods. For 25 operations in 23 patients with skull base or deep intracranial tumors (meningiomas, schwannomas, epidermoid tumors, chordomas, and others), the authors carried out presurgical simulation based on 3D CG data created by image analysis for radiological data. Interactive virtual simulation was performed by modifying the 3D CG data to imitate various surgical procedures, such as bone drilling, brain retraction, and tumor removal, with manipulation of a haptic device. The authors also produced color-printed plaster models of modified 3D CG data by a selective laser sintering method and observed them under the operative microscope.

Results. In all patients, IVS provided detailed and realistic surgical perspectives of sufficient quality, thereby allowing surgeons to determine an appropriate and feasible surgical approach. Surgeons agreed that in 44% of the 25 operations IVS showed high utility (as indicated by a rating of “prominent”) in comprehending 3D microsurgical anatomies for which reconstruction using only 2D images was complicated. Microscopic observation of color-printed plaster models in 12 patients provided further utility in confirming realistic surgical anatomies.

Conclusions. The authors’ presurgical simulation method applying advanced 3D imaging and modeling techniques provided a realistic environment for practicing microsurgical procedures virtually and enabled the authors to ascertain complex microsurgical anatomy, to determine the optimal surgical strategies, and also to efficiently educate neurosurgical trainees, especially during surgery for skull base and deep tumors.

Preoperative demonstration of the neurovascular compression characteristics with special emphasis on the degree of compression, using high-resolution magnetic resonance imaging: a prospective study, with comparison to surgical findings, in 100 consecutive patients who underwent microvascular decompression for trigeminal neuralgia

Acta Neurochir (2010) 152:817–825. DOI 10.1007/s00701-009-0588-7

Surgical outcome after microvascular decompression (MVD) for primary trigeminal neuralgia (TN) has been demonstrated as being related to the characteristics of the neurovascular compression (NVC), especially to the degree of compression exerted on the root. Therefore, preoperative determination of the NVC features could be of great value to the neurosurgeon, for evaluation of conflicting nature, exact localization, direction and degree of compression. This study deals with the predictive value of MRI in detecting and assessing features of vascular compression in 100 consecutive patients who underwent MVD for TN.

Methods. The study included 100 consecutive patients with primary TN who were submitted to a preoperative 3D MRI 1.5 T with T2 high-resolution, TOF-MRA, and T1-Gadolinium. Image analysis was performed by an independent observer blinded to the operative findings and compared with surgical data.

Findings. In 88 cases, image analysis showed NVC features that coincided with surgical findings. There were no false-positive results. Among 12 patients that did not show NVC at image analysis, nine did not have NVC at intraoperative observation, resulting in three false-negative cases. MRI sensitivity was 96.7% (88/91) and specificity 100% (9/9). Image analysis correctly identified compressible vessel in 80 of the 91 cases and degree of compression in 77 of the 91 cases. Kappa-coefficient predicting degree of root compression was 0.746, 0.767, and 0.86, respectively, for Grades I (simple contact), II (distortion), and III (marked indentation; p<0.01).

Conclusion. 3D T2 high-resolution in combination with 3D TOF-MRA and 3D T1-Gadolinium proved to be reliable in detecting NVC and in predicting the degree of the root compression