Reduction in external ventricular drain infection rate. Impact of a minimal handling protocol and antibiotic-impregnated catheters

Acta Neurochir (2011) 153:647–651. DOI 10.1007/s00701-010-0905-1

Many strategies have been developed with the aim of reducing external ventricular drain-related infections. Antibiotic-impregnated catheters are one of them.

Material and methods We report 648 cases of external ventricular drain from a total of 534 patients treated at the Virgen del Rocío Hospital between 1995 and 2006. Three subgroups were considered: group 1 included patients treated between 1995 and 2000, as well as a total of 190 external ventricular drains and 59 cases of infection (31.05%); group 2, with patients treated between 2000 and 2004 and managed with a minimal handling protocol, included 210 external ventricular drains and nine cases of infection (4.29%); and group 3, treated between 2004 and 2006, with 248 external ventricular drains and six cases of infection (2.41%). This latter subgroup included patients managed with a minimal handling protocol and antibiotic-impregnated catheters.

Results Infection rate was 17% when non-antibioticimpregnated catheters were employed and 2.41% when antibiotic-impregnated catheters were inserted (p<0.001). This difference was statistically significant before and after the introduction of a minimal handling protocol, with percentages of 5.31% and 3.27%, respectively (p<0.001; odds ratio 0.08; absolute risk reduction 27.26%). However, no statistically significant difference was observed in infection rate when the impact of a minimal handling protocol was considered: 4.29% when only the protocol was introduced and 2.41% when both the protocol and antibiotic-impregnated catheters were used (p>0.05).

Conclusion Minimal handling protocols constitute an essential strategy in the reduction of external ventricular drain-related infections. Besides that, the use of antibiotic-impregnated catheters may reduce infection-related hospital costs.

A Multicenter Multinational Registry for Assessing Ventriculoperitoneal Shunt Infections for Hydrocephalus

Neurosurgery 67:1303–1310, 2010 DOI: 10.1227/NEU.0b013e3181f07e76

Reported infection rates after ventriculoperitoneal shunt surgery vary from 1 to 25%. Antibiotic-impregnated (AI) catheters may reduce shunt infection rates, but this is uncertain.

OBJECTIVE: To establish a prospective shunt registry to evaluate short-term (3-month) infection rates associated with ventriculoperitoneal shunts and standard or AI catheters during surgical treatment of hydrocephalus.

METHODS: A prospective, multicenter, noncontrolled, open-label registry investigated patients with de novo catheter implantation or catheter replacement of an existing ventriculoperitoneal shunt. The primary outcome was shunt infection.

RESULTS: A total of 440 patients were entered into the registry at 10 sites: 3 in North America, 2 in Singapore, 4 in China and 1 in India. Seven patients were excluded. Of the 433 remaining patients, 314 had new shunts and 119 were revisions. Shunt infections occurred in 14 of 433 patients (3.2%) overall and in 2 of 37 infants (5.2%) younger than 1 year. AI catheters were used in 46 of 433 patients at 7 centers. The shunt infection rate was 0 of 46 for shunts with AI catheters and 14 of 387 (3.6%) without AI catheters. Infection rates were similar with AI catheters, adjusting for age and catheter type.

CONCLUSION: The overall shunt infection rate was lower than in previous multicentered studies. The low infection rate and low rate of AI catheter use precludes any meaningful statement regarding the value of AI catheters in reducing the infection rate. Consideration should be given to performing a well designed, adequately powered, prospective randomized controlled trial to determine whether AI catheters reduce shunt infection rates.