Freehand stereotactic ventricular catheter insertion for ventriculoperitoneal shunts based on individualized radio-anatomical landmarks

Acta Neurochirurgica (2021) 163:1103–1112

The accurate placement of the ventricular catheter (VC) is critical in reducing the incidence of proximal failure of ventriculoperitoneal shunts (VPSs). The standard freehand technique is based on validated external anatomical landmarks but remains associated with a relatively high rate of VC malposition. Already proposed alternative methods have all their specific limitations. Herein, we evaluate the accuracy of our adapted freehand technique based on an individualized radio-anatomical approach. Reproducing the preoperative imaging on the patient’s head using common anatomical landmarks allows to define stereotactic VC coordinates to be followed at surgery.

Material and methods Fifty-five consecutive patients treated with 56 VPS between 11/2005 and 02/2020 fulfilled the inclusion criteria of this retrospective study. Burr hole coordinates, VC trajectory, and length were determined in all cases on preoperative computed tomography (CT) scan and were accurately reported on patients’ head. The primary endpoint was to evaluate VC placement accuracy. The secondary endpoint was to evaluate the rate and nature of postoperative VC-related complications.

Results Our new technique was applicable in all patients and no VC-related complications were observed. Postoperative imaging showed VC optimally placed in 85.7% and sub-optimally placed in 14.3% of cases. In all procedures, all the holes on the VC tip were found in the ventricular system.

Conclusions This simple individualized technique improves the freehand VC placement in VPS surgery, making its accuracy comparable to that of more sophisticated and expensive techniques. Further randomized controlled studies are required to compare our results with those of the other available techniques.

Shunt infusion studies: impact on patient outcome, including health economics

Acta Neurochirurgica (2020) 162: 1019–1031

The diagnosis of shunt malfunction is often not straightforward. We have explored, in symptomatic shunted patients with hydrocephalus or pseudotumour cerebri syndrome (PTCS), the accuracy of CSF infusion tests in differentiating a functioning shunt from one with possible problems, and the health economic consequences.

Methods Participants: hydrocephalus/PTCS patients with infusion tests performed from January 2013 until December 2015. We followed patients up after 6 and 12 months from the test to determine whether they had improved, had persisting symptoms or had required urgent revision. We calculated the total cost savings of revision versus infusion tests and standard protocol of revision and ICP monitoring versus infusion tests.

Results Three hundred sixty-five shunt infusion tests had been performed where a shunt prechamber/reservoir was present. For hydrocephalus patients, more than half of the tests (~ 55%, 155 out of 280) showed no shunt malfunction versus 125 with possible malfunction (ages 4 months to 90 years old). For PTCS patients aged 10 to 77 years old, 47 had possible problems and 38 no indication for shunt malfunction. Overall, > 290 unnecessary revisions were avoided over 3 years’ time. Two hundred fifty eight (> 85%) of those non-surgically managed, remained well, did not deteriorate and did not require surgery. No infections were associated with infusion studies. For Cambridge, the overall savings from avoiding revisions was £945,415 annually.

Conclusions Our results provide evidence of the importance of shunt testing in vivo to confirm shunt malfunction. Avoiding unnecessary shunt revisions carries a strong health benefit for patients that also translates to a significant financial benefit for the National Health Service and potentially for other healthcare systems worldwide.