A novel miniature robotic device for frameless implantation of depth electrodes in refractory epilepsy

J Neurosurg 126:1622–1628, 2017

The authors’ group recently published a novel technique for a navigation-guided frameless stereotactic approach for the placement of depth electrodes in epilepsy patients. To improve the accuracy of the trajectory and enhance the procedural workflow, the authors implemented the iSys1 miniature robotic device in the present study into this routine.

METHODS As a first step, a preclinical phantom study was performed using a human skull model, and the accuracy and timing between 5 electrodes implanted with the manual technique and 5 with the aid of the robot were compared. After this phantom study showed an increased accuracy with robot-assisted electrode placement and confirmed the robot’s ability to maintain stability despite the rotational forces and the leverage effect from drilling and screwing, patients were enrolled and analyzed for robot-assisted depth electrode placement at the authors’ institution from January 2014 to December 2015. All procedures were performed with the S7 Surgical Navigation System with Synergy Cranial software and the iSys1 miniature robotic device.

RESULTS Ninety-three electrodes were implanted in 16 patients (median age 33 years, range 3–55 years; 9 females, 7 males). The authors saw a significant increase in accuracy compared with their manual technique, with a median deviation from the planned entry and target points of 1.3 mm (range 0.1–3.4 mm) and 1.5 mm (range 0.3–6.7 mm), respectively. For the last 5 patients (31 electrodes) of this series the authors modified their technique in placing a guide for implantation of depth electrodes (GIDE) on the bone and saw a significant further increase in the accuracy at the entry point to 1.18 ± 0.5 mm (mean ± SD) compared with 1.54 ± 0.8 mm for the first 11 patients (p = 0.021). The median length of the trajectories was 45.4 mm (range 19–102.6 mm). The mean duration of depth electrode placement from the start of trajectory alignment to fixation of the electrode was 15.7 minutes (range 8.5–26.6 minutes), which was significantly faster than with the manual technique. In 12 patients, depth electrode placement was combined with subdural electrode placement. The procedure was well tolerated in all patients. The authors did not encounter any case of hemorrhage or neurological deficit related to the electrode placement. In 1 patient with a psoriasis vulgaris, a superficial wound infection was encountered. Adequate physiological recordings were obtained from all electrodes. No additional electrodes had to be implanted because of misplacement.

CONCLUSIONS The iSys1 robotic device is a versatile and easy to use tool for frameless implantation of depth electrodes for the treatment of epilepsy. It increased the accuracy of the authors’ manual technique by 60% at the entry point and over 30% at the target. It further enhanced and expedited the authors’ procedural workflow.

Electrical stimulation of the parahippocampal gyrus for prediction of posthippocampectomy verbal memory decline

provocative-test-for-verbal-memory-decline

J Neurosurg 125:1053–1060, 2016

Epilepsy surgery is of known bene t for drug-resistant temporal lobe epilepsy (TLE); however, a certain number of patients suffer signi cant decline in verbal memory after hippocampectomy. To prevent this disabling compli- cation, a reliable test for predicting postoperative memory decline is greatly desired. Therefore, the authors assessed the value of electrical stimulation of the parahippocampal gyrus (PHG) as a provocation test of verbal memory decline after hippocampectomy on the dominant side.

Methods Eleven right-handed, Japanese-speaking patients with medically intractable left TLE participated in the study. Before surgery, they underwent provocative testing via electrical stimulation of the left PHG during a verbal en- coding task. Their pre- and posthippocampectomy memory function was evaluated according to the Wechsler Memory Scale-Revised (WMS-R) and/or Mini-Mental State Examination (MMSE) before and 6 months after surgery. The rela- tionship between postsurgical memory decline and results of the provocative test was evaluated.

Results Left hippocampectomy was performed in 7 of the 11 patients. In 3 patients with a positive provocative rec- ognition test, verbal memory function, as assessed by the WMS-R, decreased after hippocampectomy, whereas in 4 patients with a negative provocative recognition test, verbal memory function, as assessed by the WMS-R or MMSE, was preserved.

Conclusions Results of the present study suggest that electrical stimulation of the PHG is a reliable provocative test to predict posthippocampectomy verbal memory decline.