The accuracy of 3D fluoroscopy (XT) vs computed tomography (CT) registration in deep brain stimulation (DBS) surgery

Acta Neurochirurgica (2020) 162:1871–1878

Stereotactic registration is the most critical step ensuring accuracy in deep brain stimulation (DBS) surgery. 3D fluoroscopy (XT) is emerging as an alternative to CT. XT has been shown to be safe and effective for intraoperative confirmation of lead position following implantation. However, there is a lack of studies evaluating the suitability ofXT to be used for themore crucial step of registration and its capability of being merged to a preoperative MRI. This is the first study comparing accuracy, efficiency, and radiation exposure of XT- vs CT-based stereotactic registration and XT/MRI merging in deep brain stimulation.

Methods Mean absolute differences and Euclidean distance between planned (adjusted for intraoperative testing) and actual lead trajectories were calculated for accuracy of implantation. The radiation dose from each scan was recorded as the dose length product (DLP). Efficiency was measured as the time between the patient entering the operating room and the initial skin incision. A one-way ANOVA compared these parameters between patients that had either CT- or XT-based registration.

Results Forty-one patients underwent DBS surgery—25 in the CT group and 16 in the XT group. The mean absolute difference between CT and XTwas not statistically significant in the x (p = 0.331), y (p = 0.951), or z (p = 0.807) directions. The Euclidean distance between patient groups did not differ significantly (p = 0.874). The average radiation exposure with XT (220.0 ± 0.1 mGy*cm) was significantly lower than CT (1269.3 ± 112.9 mGy*cm) (p < 0.001). There was no significant difference in registration time between CT (107.8 ± 23.1 min) and XT (106.0 ± 18.2 min) (p = 0.518).

Conclusion XT-based frame registration was shown to result in similar implantation accuracy and significantly less radiation exposure compared with CT. Our results surprisingly showed no significant difference in registration time, but this may be due to a learning curve effect.

Brain Imaging in Chronic Epilepsy Patients After Depth Electrode (Stereoelectroencephalography) Implantation: Magnetic Resonance Imaging or Computed Tomography?

Depth electrode localization error

Neurosurgery 73:543–549, 2013

The accurate localization of depth electrodes in epilepsy surgery is important for correct interpretation of stereoelectroencephalography recordings and neurosurgical resection. Unfortunately, image quality in postimplantation magnetic resonance imaging (MRI) is degraded by metal artifacts. The registration of postimplantation computed tomography (CT) or MRI to preimplantation (artifact-free) MRI facilitates electrode imaging and optimal visualization of brain anatomy. However, registration errors negatively affect electrode localization accuracy.

OBJECTIVE: To compare the relative registration deviation between postimplantation CT and MRI with preimplantation MRI.

METHODS: Retrospectively, 14 pharmacoresistant epilepsy patients were included who underwent stereotactic insertion of multiple depth electrodes and preimplantation and postimplantation MRI and postimplantation CT. Postimplantation MRI and CT image sets were registered to preimplantation MRI. The registration error between the registered postimplantation MRI and CT was quantified by measuring the geometrical distance between the electrodes of the registered postimplantation CT and the postimplantation MRI.

RESULTS: The registration error of postimplantation imaging to preimplantation MRI was dependent on the algorithm used. After optimization, the smallest registration error was 1.22 6 0.29 mm (mean 6 SD) at the tip and 2.25 6 1.18 mm at the base of the electrode.

CONCLUSION: The good correspondence between the CT/MRI and the MRI/MRI registration suggests that either postimplantation MRI or CT is sufficient for accurate electrode localization. In case of postoperative morphological brain deformations, postimplantation MRI is still recommended.

Is MRI a reliable tool to locate the electrode after deep brain stimulation surgery? Comparison study of CT and MRI for the localization of electrodes after DBS

Acta Neurochir (2010) 152:2029–2036. DOI 10.1007/s00701-010-0779-2

MRI has been utilized to localize the electrode after deep brain stimulation, but its accuracy has been questioned due to image distortion. Under the hypothesis that MRI is not adequate for evaluation of electrode position after deep brain stimulation, this study is aimed at validating the accuracy of MRI in electrode localization in comparison with CT scan. Methods Sixty one patients who had undergone STN DBS were enrolled for the analysis. Using mutual information technique, CT and MRI taken at 6 months after the operation were fused. The x and y coordinates of the centers of electrodes shown of CT and MRI were compared in the fused images to calculate average difference at five different levels. The difference of the tips of the electrodes, designated as the z coordinate, was also calculated. Results The average of the distance between the centers of the electrodes in the five levels estimated in the fused image of brain CT and MRI taken at least 6 months after STN DBS was 1.33 mm (0.1–5.8 mm). The average discrepancy of x coordinates for all five levels between MRI and CT was 0.56±0.54 mm (0–5.7 mm), the discrepancy of y coordinates was 1.06±0.59 mm (0–3.5 mm), and for the z coordinate, it was 0.98±0.52 mm (0–3.1 mm) (all p values <0.001). Notably, the average discrepancy of x coordinates at 3.5 mm below AC–PC level, i.e., at the STN level between MRI and CT, was 0.59±0.42 mm (0–2.4 mm); the discrepancy of y coordinates was 0.81±0.47 mm (0–2.9 mm) (p values<0.001). Conclusions The results suggest that there was significant discrepancy between the centers of electrodes estimated by CT and MRI after STN DBS surgery.