Intraoperative computed tomography in epilepsy surgery

Intraoperative computed tomography for intracranial electrode implantation surgery in medically refractory epilepsy

J Neurosurg 122:526–531, 2015

Accurate placement of intracranial depth and subdural electrodes is important in evaluating patients with medically refractory epilepsy for possible resection. Confirming electrode locations on postoperative CT scans does not allow for immediate replacement of malpositioned electrodes, and thus revision surgery is required in select cases. Intraoperative CT (iCT) using the Medtronic O-arm device has been performed to detect electrode locations in deep brain stimulation surgery, but its application in epilepsy surgery has not been explored. In the present study, the authors describe their institutional experience in using the O-arm to facilitate accurate placement of intracranial electrodes for epilepsy monitoring.

Methods In this retrospective study, the authors evaluated consecutive patients who had undergone subdural and/or depth electrode implantation for epilepsy monitoring between November 2010 and September 2012. The O-arm device is used to obtain iCT images, which are then merged with the preoperative planning MRI studies and reviewed by the surgical team to confirm final positioning. Minor modifications in patient positioning and operative field preparation are necessary to safely incorporate the O-arm device into routine intracranial electrode implantation surgery. The device does not obstruct surgeon access for bur hole or craniotomy surgery. Depth and subdural electrode locations are easily identified on iCT, which merge with MRI studies without difficulty, allowing the epilepsy surgical team to intraoperatively confirm lead locations.

Results Depth and subdural electrodes were implanted in 10 consecutive patients by using routine surgical techniques together with preoperative stereotactic planning and intraoperative neuronavigation. No wound infections or other surgical complications occurred. In one patient, the hippocampal depth electrode was believed to be in a suboptimal position and was repositioned before final wound closure. Additionally, 4 strip electrodes were replaced due to suboptimal positioning. Postoperative CT scans did not differ from iCT studies in the first 3 patients in the series and thus were not obtained in the final 7 patients. Overall, operative time was extended by approximately 10–15 minutes for O-arm positioning, less than 1 minute for image acquisition, and approximately 10 minutes for image transfer, fusion, and intraoperative analysis (total time 21–26 minutes).

Conclusions The O-arm device can be easily incorporated into routine intracranial electrode implantation surgery in standard-sized operating rooms. The technique provides accurate 3D visualization of depth and subdural electrode contacts, and the intraoperative images can be easily merged with preoperative MRI studies to confirm lead positions before final wound closure. Intraoperative CT obviates the need for routine postoperative CT and has the potential to improve the accuracy of intracranial electroencephalography recordings and may reduce the necessity for revision surgery.

The stereotactic approach for mapping epileptic networks

pb-red28

J Neurosurg 121:1239–1246, 2014

Stereoelectroencephalography (SEEG) is a methodology that permits accurate 3D in vivo electroclinical recordings of epileptiform activity. Among other general indications for invasive intracranial electroencephalography (EEG) monitoring, its advantages include access to deep cortical structures, its ability to localize the epileptogenic zone when subdural grids have failed to do so, and its utility in the context of possible multifocal seizure onsets with the need for bihemispheric explorations. In this context, the authors present a brief historical overview of the technique and report on their experience with 2 SEEG techniques (conventional Leksell frame-based stereotaxy and frameless stereotaxy under robotic guidance) for the purpose of invasively monitoring difficult-to-localize refractory focal epilepsy.

Methods. Over a period of 4 years, the authors prospectively identified 200 patients with refractory epilepsy who collectively underwent 2663 tailored SEEG electrode implantations for invasive intracranial EEG monitoring and extraoperative mapping. The first 122 patients underwent conventional Leksell frame-based SEEG electrode placement; the remaining 78 patients underwent frameless stereotaxy under robotic guidance, following acquisition of a stereotactic ROSA robotic device at the authors’ institution. Electrodes were placed according to a preimplantation hypothesis of the presumed epileptogenic zone, based on a standardized preoperative workup including video-EEG monitoring, MRI, PET, ictal SPECT, and neuropsychological assessment. Demographic features, seizure semiology, number and location of implanted SEEG electrodes, and location of the epileptogenic zone were recorded and analyzed for all patients. For patients undergoing subsequent craniotomy for resection, the type of resection and procedure-related complications were prospectively recorded. These results were analyzed and correlated with pathological diagnosis and postoperative seizure outcomes.

Results. The epileptogenic zone was confirmed by SEEG in 154 patients (77%), of which 134 (87%) underwent subsequent craniotomy for epileptogenic zone resection. Within this cohort, 90 patients had a minimum follow-up of at least 12 months; therein, 61 patients (67.8%) remained seizure free, with an average follow-up period of 2.4 years. The most common pathological diagnosis was focal cortical dysplasia Type I (55 patients, 61.1%). Per electrode, the surgical complications included wound infection (0.08%), hemorrhagic complications (0.08%), and a transient neurological deficit (0.04%) in a total of 5 patients (2.5%). One patient (0.5%) ultimately died due to intracerebral hematoma directly ensuing from SEEG electrode placement.

Conclusions. Based on these results, SEEG methodology is safe, reliable, and effective. It is associated with minimal morbidity and mortality, and serves as a practical, minimally invasive approach to extraoperative localization of the epileptogenic zone in patients with refractory epilepsy.

Surgically Relevant Localization of the Central Sulcus With High-Density Somatosensory-Evoked Potentials Compared With Functional Magnetic Resonance Imaging

Surgically Relevant Localization of the Central Sulcus With High-Density Somatosensory-Evoked Potentials Compared With Functional Magnetic Resonance Imaging

Neurosurgery 74:517–526, 2014

Resection of abnormal brain tissue lying near the sensorimotor cortex entails precise localization of the central sulcus. Mapping of this area is achieved by applying invasive direct cortical electrical stimulation. However, noninvasive methods, particularly functional magnetic resonance imaging (fMRI), are also used. As a supplement to fMRI, localization of somatosensory-evoked potentials (SEPs) recorded with an electroencephalogram (EEG) has been proposed, but has not found its place in clinical practice.

OBJECTIVE: To assess localization accuracy of the hand somatosensory cortex with SEP source imaging.

METHODS: We applied electrical source imaging in 49 subjects, recorded with highdensity EEG (256 channels). We compared it with fMRI in 18 participants and with direct cortical electrical stimulation in 6 epileptic patients.

RESULTS: Comparison of SEP source imaging with fMRI indicated differences of 3 to 8 mm, with the exception of the mesial-distal orientation, where variances of up to 20 mm were found. This discrepancy is explained by the fact that the source maximum of the first SEP peak is localized deep in the central sulcus (area 3b), where information initially arrives. Conversely, fMRI showed maximal signal change on the lateral surface of the postcentral gyrus (area 1), where sensory information is integrated later in time. Electrical source imaging and fMRI showed mean Euclidean distances of 13 and 14 mm, respectively, from the contacts where electrocorticography elicited sensory phenomena of the contralateral upper limb.

CONCLUSION: SEP source imaging, based on high-density EEG, reliably identifies the depth of the central sulcus. Moreover, it is a simple, flexible, and relatively inexpensive alternative to fMRI.

Stereotactic electroencephalography with temporal grid and mesial temporal depth electrode coverage: does technique of depth electrode placement affect outcome?

J Neurosurg 113:32–38, 2010.DOI: 10.3171/2009.12.JNS091073

Intracranial monitoring for temporal lobe seizure localization to differentiate neocortical from mesial temporal onset seizures requires both neocortical subdural grids and hippocampal depth electrode implantation. There are 2 basic techniques for hippocampal depth electrode implantation. This first technique uses a stereotactically guided 8-contact depth electrode directed along the long axis of the hippocampus to the amygdala via an occipital bur hole. The second technique involves direct placement of 2 or 3 4-contact depth electrodes perpendicular to the temporal lobe through the middle temporal gyrus and overlying subdural grid. The purpose of this study was to determine whether one technique was superior to the other by examining monitoring success and complications.

Methods. Between 1997 and 2005, 41 patients underwent invasive seizure monitoring with both temporal subdural grids and depth electrodes placed in 2 ways. Patients in Group A underwent the first technique, and patients in Group B underwent the second technique.

Results. Group A consisted of 26 patients and Group B 15 patients. There were no statistically significant differences between Groups A and B regarding demographics, monitoring duration, seizure localization, or outcome (Engel classification). There was a statistically significant difference at the point in time at which these techniques were used: Group A represented more patients earlier in the series than Group B (p < 0.05). The complication rate attributable to the grids and depth electrodes was 0% in each group. It was more likely that the depth electrodes were placed through the grid if there was a prior resection and the patient was undergoing a new evaluation (p < 0.05). Furthermore, Group A procedures took significantly longer than Group B procedures.

Conclusions. In this patient series, there was no difference in efficacy of monitoring, complications, or outcome between hippocampal depth electrodes placed laterally through temporal grids or using an occipital bur hole stereotactic approach. Placement of the depth electrodes perpendicularly through the grids and middle temporal gyrus is technically more practical because multiple head positions and redraping are unnecessary, resulting in shorter operative times with comparable results.