The effect of augmented reality on the accuracy and learning curve of external ventricular drain placement

Neurosurg Focus 51 (2):E8, 2021

The traditional freehand technique for external ventricular drain (EVD) placement is most frequently used, but remains the primary risk factor for inaccurate drain placement. As this procedure could benefit from image guidance, the authors set forth to demonstrate the impact of augmented-reality (AR) assistance on the accuracy and learning curve of EVD placement compared with the freehand technique.

METHODS Sixteen medical students performed a total of 128 EVD placements on a custom-made phantom head, both before and after receiving a standardized training session. They were guided by either the freehand technique or by AR, which provided an anatomical overlay and tailored guidance for EVD placement through inside-out infrared tracking. The outcome was quantified by the metric accuracy of EVD placement as well as by its clinical quality.

RESULTS The mean target error was significantly impacted by either AR (p = 0.003) or training (p = 0.02) in a direct comparison with the untrained freehand performance. Both untrained (11.9 ± 4.5 mm) and trained (12.2 ± 4.7 mm) AR performances were significantly better than the untrained freehand performance (19.9 ± 4.2 mm), which improved after training (13.5 ± 4.7 mm). The quality of EVD placement as assessed by the modified Kakarla scale (mKS) was significantly impacted by AR guidance (p = 0.005) but not by training (p = 0.07). Both untrained and trained AR performances (59.4% mKS grade 1 for both) were significantly better than the untrained freehand performance (25.0% mKS grade 1). Spatial aptitude testing revealed a correlation between perceptual ability and untrained AR-guided performance (r = 0.63).

CONCLUSIONS Compared with the freehand technique, AR guidance for EVD placement yielded a higher outcome accuracy and quality for procedure novices. With AR, untrained individuals performed as well as trained individuals, which indicates that AR guidance not only improved performance but also positively impacted the learning curve. Future efforts will focus on the translation and evaluation of AR for EVD placement in the clinical setting.

Image guidance and improved accuracy of external ventricular drain tip position particularly in patients with small ventricles

J Neurosurg 130:1268–1273, 2019

External ventricular drain (EVD) insertion is one of the most common emergency neurosurgical procedures. EVDs are traditionally inserted freehand (FH) in an emergency setting, but often result in suboptimal positioning. Image-guided surgery (IGS) is selectively used to assist placement. However, the accuracy and practicality of IGS use is yet to be reported. In this study, the authors set out to assess if IGS is practical and improves the accuracy of EVD placement.

METHODS Case notes and images obtained in patients who underwent frontal EVD placement were retrospectively reviewed. Ventriculomegaly was determined by the measurement of the Evans index. EVD location was classified as optimal (ipsilateral frontal horn) or suboptimal (any other location). Propensity score matching of the two groups (IGS vs FH) for the Evans index was performed. Data were analyzed for patient age, diagnosis, number of EVDs, and complications. Those without postoperative CT scans were excluded.

RESULTS A total of 607 patients with 760 EVDs placed were identified; 331 met inclusion criteria. Of these, 287 were inserted FH, and 44 were placed with IGS; 60.6% of all unmatched FH EVDs were optimal compared with 75% of the IGS group (p = 0.067). The IGS group had a significantly smaller Evans index (p < 0.0001). Propensity score matching demonstrated improved optimal position in the IGS group when compared with the matched FH group (75% vs 43.2%, OR 4.6 [1.5–14.6]; p = 0.002). Patients with an Evans index of ≥ 0.36 derived less benefit (75% in IGS vs 66% in FH, p = 0.5), and those with an Evans index < 0.36 derived more benefit (75% in IGS vs 53% in FH, p = 0.024). The overall EVD complication rate was 36% in the FH group versus 18% in the IGS group (p = 0.056). Revision rates were higher in the FH group (p = 0.035), and the operative times were similar (p = 0.69). Long intracranial EVD catheters were associated with tip malposition irrespective of the group.

CONCLUSIONS Image guidance is practical and improves the accuracy of EVD placement in patients with small ventricles; thus, it should be considered for these patients.