Inducing transient language disruptions by mapping of Broca’s area with modified patterned repetitive transcranial magnetic stimulation protocol

Mapping Broca’s area

J Neurosurg 120:1033–1041, 2014

Until now there has been no reliable stimulation protocol for inducing transient language disruptions while mapping Broca’s area. Despite the promising data of only a few studies in which speech arrest and language disturbances have been induced, certain concerns have been raised. The purpose of this study was to map Broca’s area by using event-related navigated transcranial magnetic stimulation (nTMS) to generate a modified patterned nTMS protocol.

Methods. Eleven right-handed subjects underwent nTMS to Broca’s area while engaged in a visual object-naming task. Navigated TMS was triggered 300 msec after picture presentation. The modified patterned nTMS protocol consists of 4 stimuli with an interstimulus interval of 6 msec; 8 or 16 of those bursts were repeated with a burst repetition rate of 12 Hz. Prior to mapping of Broca’s area, the primary motor cortices (M1) for hand and laryngeal muscles were mapped. The Euclidian distance on MRI was measured between cortical points eliciting transient language disruptions and M1 for the laryngeal muscle.

Results. On stimulating Broca’s area, transient language disruptions were induced in all subjects. The mean Euclidian distance between cortical spots inducing transient language disruptions and M1 for the laryngeal muscle was 17.23 ± 4.73 mm.

Conclusions. The stimulation paradigm with the modified patterned nTMS protocol was shown to be promising and might gain more widespread use in speech localization in clinical and research applications.

Continuous dynamic mapping of the corticospinal tract during surgery of motor eloquent brain tumors

Continuous dynamic subcortical mapping

J Neurosurg 120:1015–1024, 2014

The authors developed a new mapping technique to overcome the temporal and spatial limitations of classic subcortical mapping of the corticospinal tract (CST). The feasibility and safety of continuous (0.4–2 Hz) and dynamic (at the site of and synchronized with tissue resection) subcortical motor mapping was evaluated.

Methods. The authors prospectively studied 69 patients who underwent tumor surgery adjacent to the CST (< 1 cm using diffusion tensor imaging and fiber tracking) with simultaneous subcortical monopolar motor mapping (short train, interstimulus interval 4 msec, pulse duration 500 μsec) and a new acoustic motor evoked potential alarm. Continuous (temporal coverage) and dynamic (spatial coverage) mapping was technically realized by integrating the mapping probe at the tip of a new suction device, with the concept that this device will be in contact with the tissue where the resection is performed. Motor function was assessed 1 day after surgery, at discharge, and at 3 months.

Results. All procedures were technically successful. There was a 1:1 correlation of motor thresholds for stimulation sites simultaneously mapped with the new suction mapping device and the classic fingerstick probe (24 patients, 74 stimulation points; r2 = 0.98, p < 0.001). The lowest individual motor thresholds were as follows: > 20 mA, 7 patients; 11–20 mA, 13 patients; 6–10 mA, 8 patients; 4–5 mA, 17 patients; and 1–3 mA, 24 patients. At 3 months, 2 patients (3%) had a persistent postoperative motor deficit, both of which were caused by a vascular injury. No patient had a permanent motor deficit caused by a mechanical injury of the CST.

Conclusions. Continuous dynamic mapping was found to be a feasible and ergonomic technique for localizing the exact site of the CST and distance to the motor fibers. The acoustic feedback and the ability to stimulate the tissue continuously and exactly at the site of tissue removal improves the accuracy of mapping, especially at low (< 5 mA) stimulation intensities. This new technique may increase the safety of motor eloquent tumor surgery.

Fiber tractography of basal ganglia and cerebellum in PD

Fiber tractography of basal ganglia and cerebellum in PD

J Neurosurg 120:988–996, 2014

Stimulation of white matter pathways near targeted structures may contribute to therapeutic effects of deep brain stimulation (DBS) for patients with Parkinson disease (PD). Two tracts linking the basal ganglia and cerebellum have been described in primates: the subthalamopontocerebellar tract (SPCT) and the dentatothalamic tract (DTT). The authors used fiber tractography to evaluate white matter tracts that connect the cerebellum to the region of the basal ganglia in patients with PD who were candidates for DBS.

Methods. Fourteen patients with advanced PD underwent 3-T MRI, including 30-directional diffusion-weighted imaging sequences. Diffusion tensor tractography was performed using 2 regions of interest: ipsilateral subthalamic and red nuclei, and contralateral cerebellar hemisphere. Nine patients underwent subthalamic DBS, and the course of each tract was observed relative to the location of the most effective stimulation contact and the volume of tissue activated.

Results. In all patients 2 distinct tracts were identified that corresponded closely to the described anatomical features of the SPCT and DTT, respectively. The mean overall distance from the active contact to the DTT was 2.18 ± 0.35 mm, and the mean proportional distance relative to the volume of tissue activated was 1.35 ± 0.48. There was a nonsignificant trend toward better postoperative tremor control in patients with electrodes closer to the DTT.

Conclusions. The SPCT and the DTT may be related to the expression of symptoms in PD, and this may have implications for DBS targeting. The use of tractography to identify the DTT might assist with DBS targeting in the future.

Intraoperative optical imaging for functional mapping

Intraoperative optical imaging for functional mapping

J Neurosurg 119:853–863, 2013

Intraoperative optical imaging (IOI) is an experimental technique used for visualizing functional brain areas after surgical exposure of the cerebral cortex. This technique identifies areas of local changes in blood volume and oxygenation caused by stimulation of specific brain functions. The authors describe a new IOI method, including innovative data analysis, that can facilitate intraoperative functional imaging on a routine basis. To evaluate the reliability and validity of this approach, they used the new IOI method to demonstrate visualization of the median nerve area of the somatosensory cortex.

Methods. In 41 patients with tumor lesions adjacent to the postcentral gyrus, lesions were surgically removed by using IOI during stimulation of the contralateral median nerve. Optical properties of the cortical tissue were measured with a sensitive camera system connected to a surgical microscope. Imaging was performed by using 9 cycles of alternating prolonged stimulation and rest periods of 30 seconds. Intraoperative optical imaging was based on blood volume changes detected by using a filter at an isosbestic wavelength (delta = 568 nm). A spectral analysis algorithm was used to improve computation of the activity maps. Movement artifacts were compensated for by an elastic registration algorithm. For validation, intraoperative conduction of the phase reversal over the central sulcus and postoperative evaluation of the craniotomy site were used.

Results. The new method and analysis enabled significant differentiation (p < 0.005) between functional and nonfunctional tissue. The identification and visualization of functionally intact somatosensory cortex was highly reliable; sensitivity was 94.4% and specificity was almost 100%. The surgeon was provided with a 2D high-resolution activity map within 12 minutes. No method-related side effects occurred in any of the 41 patients.

Conclusions. The authors’ new approach makes IOI a contact-free and label-free optical technique that can be used safely in a routine clinical setup. Intraoperative optical imaging can be used as an alternative to other methods for the identification of sensory cortex areas and offers the added benefit of a high-resolution map of functional activity. It has great potential for visualizing and monitoring additional specific functional brain areas such as the visual, motor, and speech cortex. A prospective national multicenter clinical trial is currently being planned.

Accuracy of deep brain stimulation electrode placement using intraoperative computed tomography without microelectrode recording

iCT DBS

J Neurosurg 119:301–306, 2013

In this prospective study the authors’ objective was to evaluate the accuracy of deep brain stimulation (DBS) electrode placement using image guidance for direct anatomical targeting with intraoperative CT.

Methods. Preoperative 3-T MR images were merged with intraoperative CT images for planning. Electrode targets were anatomical, based on the MR images. A skull-mounted NexFrame system was used for electrode placement, and all procedures were performed under general anesthesia. After electrode placement, intraoperative CT images were merged with trajectory planning images to calculate accuracy. Accuracy was assessed by both vector error and deviation off the planned trajectory.

Results. Sixty patients (33 with Parkinson disease, 26 with essential tremor, and 1 with dystonia) underwent the procedure. Patient’s mean age was 64 ± 9.5 years. Over an 18-month period, 119 electrodes were placed (all bilateral, except one). Electrode implant locations were the ventral intermediate nucleus (VIM), globus pallidus internus (GPI), and subthalamic nucleus (STN) in 25, 23, and 12 patients, respectively. Target accuracy measurements were as follows: mean vector error 1.59 ± 1.11 mm and mean deviation off trajectory 1.24 ± 0.87 mm. There was no statistically significant difference between the accuracy of left and right brain electrodes. There was a statistically significant (negative) correlation between the distance of the closest approach of the electrode trajectory to the ventricular wall of the lateral ventricle and vector error (r2 = -0.339, p < 0.05, n = 76), and the deviation from the planned trajectory (r2 = -0.325, p < 0.05, n = 77). Furthermore, when the distance from the electrode trajectory and the ventricular wall was < 4 mm, the correlation of the ventricular distance to the deviation from the planned trajectory was stronger (r2 = -0.419, p = 0.05, n = 19). Electrodes placed in the GPI were significantly more accurate than those placed in the VIM (p < 0.05). Only 1 of 119 electrodes required intraoperative replacement due to a vector error > 3 mm. In this series there was one infection and no intraparenchymal hemorrhages.

Conclusions. Placement of DBS electrodes using an intraoperative CT scanner and the NexFrame achieves an accuracy that is at least comparable to other methods.

Long-term results of posteromedial hypothalamic deep brain stimulation for patients with resistant aggressiveness

Posteromedial hypothalamic DBS for intractable erethism

J Neurosurg 119:277–287, 2013

Erethism describes severe cases of unprovoked aggressive behavior, usually associated with some degree of mental impairment and gross brain damage. The etiology can be epileptic, postencephalitic, or posttraumatic, or the condition can be caused by brain malformations or perinatal insults. Erethism is often refractory to medication, and patients must often be interned in institutions, where they are managed with major restraining measures. The hypothalamus is a crucial group of nuclei that coordinate behavioral and autonomic responses and play a central role in the control of aggressive behavior. Deep brain stimulation (DBS) of the posteromedial hypothalamus (PMH) has been proposed as a treatment for resistant erethism, although experience with this treatment around the world is scarce. The objective of this study was to examine the long-term outcome of PMH DBS in 6 patients with severe erethism treated at the authors’ institution.

Methods. Medical records of 6 patients treated with PMH DBS for intractable aggressiveness were reviewed. The therapeutic effect on behavior was assessed by the Inventory for Client and Agency Planning preoperatively and at the last follow-up visit.

Results. Two patients died during the follow-up period due to causes unrelated to the neurosurgical treatment. Five of 6 patients experienced a significant reduction in aggressiveness (the mean Inventory for Client and Agency Planning general aggressiveness score was -47 at baseline and -25 at the last follow-up; mean follow-up 3.5 years). Similar responses were obtained with low- and high-frequency stimulation. In 4 cases, the patients’ sleep patterns became more regular, and in 1 case, binge eating and polydipsia ceased. One of the 3 patients who had epilepsy noticed a 30% reduction in seizure frequency. Another patient experienced a marked sympathetic response with highfrequency stimulation during the first stimulation trial, but this subsided when stimulation was set at low frequency. A worsening of a previous headache was noted by 1 patient. There were no other side effects.

Conclusions. In this case series, 5 of 6 patients with pathological aggressiveness had a reduction of their outbursts of violence after PMH DBS, without significant adverse effects. Prospective controlled studies with a larger number of patients are needed to confirm these results.

Comparison of the accuracy of ventricular catheter placement using freehand placement, ultrasonic guidance, and stereotactic neuronavigation

Stereotactic Ventricular Catheter Placement

J Neurosurg 119:66–70, 2013

The objective of this study was to compare the accuracy of 3 methods of ventricular catheter placement during CSF shunt operations: the freehand technique using surface anatomy, ultrasonic guidance, and stereotactic neuronavigation.

Methods. This retrospective cohort study included all patients from a single institution who underwent a ventricular CSF shunting procedure in which a new ventricular catheter was placed between January 2005 and March 2010. Data abstracted for each patient included age, sex, diagnosis, method of ventricular catheter placement, site and side of ventricular catheter placement, Evans ratio, and bifrontal ventricular span. Postoperative radiographic studies were reviewed for accuracy of ventricular catheter placement. Medical records were also reviewed for evidence of shunt failure requiring revision through December 2011. Statistical analysis was then performed comparing the 3 methods of ventricular catheter placement and to determine risk factors for inaccurate placement.

Results. There were 249 patients included in the study; 170 ventricular catheters were freehand passed, 51 were placed using stereotactic neuronavigation, and 28 were placed under intraoperative ultrasonic guidance. There was a statistically significant difference between freehand catheters and stereotactic-guided catheters (p < 0.001), as well as between freehand catheters and ultrasound-guided catheters (p < 0.001). The only risk factor for inaccurate placement identified in this study was use of the freehand technique. The use of stereotactic neuronavigation and ultrasonic guidance reduced proximal shunt failure rates (p < 0.05) in comparison with a freehand technique.

Conclusions. Stereotactic- and ultrasound-guided ventricular catheter placements are significantly more accurate than freehand placement, and the use of these intraoperative guidance techniques reduced proximal shunt failure in this study.

Endoscopic endonasal skull base surgery in the pediatric population

Epeds12160-1

J Neurosurg Pediatrics 11:227–241, 2013

The use of endoscopic endonasal surgery (EES) for skull base pathologies in the pediatric population presents unique challenges and has not been well described. The authors reviewed their experience with endoscopic endonasal approaches in pediatric skull base surgery to assess surgical outcomes and complications in the context of presenting patient demographics and pathologies.

Methods. A retrospective review of 133 pediatric patients who underwent EES at our institution from July 1999 to May 2011 was performed.

Results. A total of 171 EESs were performed for skull base tumors in 112 patients and bony lesions in 21. Eightyfive patients (63.9%) were male, and the mean age at the time of surgery was 12.7 years (range 2.3–18.0 years). Skull base tumors included angiofibromas (n = 24), craniopharyngiomas (n = 16), Rathke cleft cysts (n = 12), pituitary adenomas (n = 11), chordomas/chondrosarcomas (n = 10), dermoid/epidermoid tumors (n = 9), and 30 other pathologies. In total, 19 tumors were malignant (17.0%). Among patients with follow-up data, gross-total resection was achieved in 16 cases of angiofibromas (76.2%), 9 of craniopharyngiomas (56.2%), 8 of Rathke cleft cysts (72.7%), 7 of pituitary adenomas (70%), 5 of chordomas/chondrosarcomas (50%), 6 of dermoid/epidermoid tumors (85.7%), and 9 cases of other pathologies (31%). Fourteen patients received adjuvant radiotherapy, and 5 received chemotherapy. Sixteen patients (15.4%) showed tumor recurrence and underwent reoperation. Bony abnormalities included skull base defects (n = 12), basilar invagination (n = 4), optic nerve compression (n = 3) and trauma (n = 2); preexisting neurological dysfunction resolved in 12 patients (57.1%), improved in 7 (33.3%), and remained unchanged in 2 (9.5%). Overall, complications included CSF leak in 14 cases (10.5%), meningitis in 5 (3.8%), transient diabetes insipidus in 8 patients (6.0%), and permanent diabetes insipidus in 12 (9.0%). Five patients (3.8%) had transient and 3 (2.3%) had permanent cranial nerve palsies. The mean follow-up time was 22.7 months (range 1–122 months); 5 patients were lost to follow-up.

Conclusions. Endoscopic endonasal surgery has proved to be a safe and feasible approach for the management of a variety of pediatric skull base pathologies. When appropriately indicated, EES may achieve optimal outcomes in the pediatric population

Limbic system surgery for treatment-refractory obsessive-compulsive disorder

Limbic system surgery for treatment-refractory obsessive-compulsive disorder

J Neurosurg 118:491–497, 2013

Obsessive-compulsive disorder (OCD) is a common and disabling psychiatric illness, and in a significant proportion of patients with OCD the disease is refractory to conventional pharmacotherapy and psychotherapy. For more than half a century, patients with severe, treatment-resistant OCD have been treated with stereotactic limbic system lesions, including dorsal anterior cingulotomy. The authors present their results describing the efficacy and durability of limbic system surgery for OCD, characterizing a large cohort of patients treated at a single institution with a mean follow-up of more than 5 years.

Methods. The authors identified 64 consecutive patients undergoing cingulotomy for refractory OCD at the Massachusetts General Hospital between 1989 and 2009. Changes in OCD and major depressive disorder symptom severity were assessed at both the initial and most recent postoperative follow-up by using the Yale-Brown Obsessive Compulsive Scale and the Beck Depression Inventory, respectively. Full and partial OCD symptom responses were defined as Yale-Brown Obsessive Compulsive Scale score reductions of ≥ 35% and 25%–34%, respectively.

Results. Regarding OCD symptom improvement, at the first postoperative follow-up (mean 10.7 months), 35% of patients demonstrated a full response and 7% were partial responders. Thirty patients had a subsequent procedure (repeat cingulotomy or subcaudate tractotomy). By the most recent follow-up (mean 63.8 months), rates climbed to 47% and 22% for full and partial responses, respectively. Of the 24 patients with at least a partial response at initial follow-up, 20 (83%) retained at least a partial response at final follow-up. Comorbid major depressive disorder severity decreased by 17% at the most recent follow-up.

Conclusions. Limbic system surgery based on initial cingulotomy offers a durable and effective treatment option for appropriately selected patients with severe OCD who have not responded to conventional pharmacotherapy or psychotherapy.

Limbic system surgery for treatment-refractory obsessive-compulsive disorder

Limbic system surgery for treatment-refractory obsessive-compulsive disorder- a prospective long-term follow-up of 64 patients

DOI: 10.3171/2012.11.JNS12389

Obsessive-compulsive disorder (OCD) is a common and disabling psychiatric illness, and in a significant proportion of patients with OCD the disease is refractory to conventional pharmacotherapy and psychotherapy. For more than half a century, patients with severe, treatment-resistant OCD have been treated with stereotactic limbic system lesions, including dorsal anterior cingulotomy.

The authors present their results describing the efficacy and durability of limbic system surgery for OCD, characterizing a large cohort of patients treated at a single institution with a mean follow-up of more than 5 years.

Methods. The authors identified 64 consecutive patients undergoing cingulotomy for refractory OCD at the Massachusetts General Hospital between 1989 and 2009. Changes in OCD and major depressive disorder symptom severity were assessed at both the initial and most recent postoperative follow-up by using the Yale-Brown Obsessive Compulsive Scale and the Beck Depression Inventory, respectively. Full and partial OCD symptom responses were defined as Yale-Brown Obsessive Compulsive Scale score reductions of ≥ 35% and 25%–34%, respectively.

Results. Regarding OCD symptom improvement, at the first postoperative follow-up (mean 10.7 months), 35% of patients demonstrated a full response and 7% were partial responders. Thirty patients had a subsequent procedure (repeat cingulotomy or subcaudate tractotomy). By the most recent follow-up (mean 63.8 months), rates climbed to 47% and 22% for full and partial responses, respectively. Of the 24 patients with at least a partial response at initial follow-up, 20 (83%) retained at least a partial response at final follow-up. Comorbid major depressive disorder severity decreased by 17% at the most recent follow-up.

Conclusions. Limbic system surgery based on initial cingulotomy offers a durable and effective treatment option for appropriately selected patients with severe OCD who have not responded to conventional pharmacotherapy or psychotherapy.

Analysis of lumbar plexopathies and nerve injury after lateral retroperitoneal transpsoas approach

Analysis of lumbar plexopathies and nerve injury after lateral retroperitoneal transpsoas approach

DOI: 10.3171/2012.11.SPINE12755

The minimally invasive lateral transpsoas approach has become an increasingly popular means of fusion. The most frequent complication is related to lumbar plexus nerve injuries; these can be diagnosed based on distribution of neurological deficit following the motor and/or sensory nerve injury. However, the literature has failed to provide a clinically relevant description of these complications. With accurate clinical diagnosis, spine practitioners can provide more precise prognostic and management recommendations to include observation, nerve blocks, neurodestructive procedures, medications, or surgical repair strategies. The purpose of this study was to standardize the clinical findings of lumbar plexopathies and nerve injuries associated with minimally invasive lateral retroperitoneal transpsoas lumbar fusion.

Methods. A thorough literature search of the MEDLINE database up to June 2012 was performed to identify studies that reported lumbar plexus and nerve injuries after the minimally invasive lateral retroperitoneal transpsoas approach. Included studies were assessed for described neurological deficits postoperatively. Studies that did attempt to describe nerve-related complications clinically were excluded. A clinically relevant assessment of lumbar plexus nerve injury was derived to standardize early diagnosis and outline prognostic implications.

Results. A total of 18 studies were selected with a total of 2310 patients; 304 patients were reported to have possible plexus-related complications. The incidence of documented nerve and/or root injury and abdominal paresis ranged from 0% to 3.4% and 4.2%, respectively. Motor weakness ranged from 0.7% to 33.6%. Sensory complications ranged from 0% to 75%. A lack of consistency in the descriptions of the lumbar plexopathies and/or nerve injuries as well as a lack of diagnostic paradigms was noted across studies reviewed. Sensory dermal zones were established and a standardized approach was proposed.

Conclusions. There is underreporting of postoperative lumbar plexus nerve injury and a lack of standardization of clinical findings of neural complications related to the minimally invasive lateral retroperitoneal transpsoas approach. The authors provide a diagnostic paradigm that allows for an efficient and accurate classification of postoperative lumbar plexopathies and nerve injuries.

Short circuit in deep brain stimulation

J Neurosurg 117:955–961, 2012

The authors undertook this study to investigate the incidence, cause, and clinical influence of short circuits in patients treated with deep brain stimulation (DBS).

Methods. After the incidental identification of a short circuit during routine follow-up, the authors initiated a policy at their institution of routinely evaluating both therapeutic impedance and system impendence at every outpatient DBS follow-up visit, irrespective of the presence of symptoms suggesting possible system malfunction. This study represents a report of their findings after 1 year of this policy.

Results. Implanted DBS leads exhibiting short circuits were identified in 7 patients (8.9% of the patients seen for outpatient follow-up examinations during the 12-month study period). The mean duration from DBS lead implantation to the discovery of the short circuit was 64.7 months. The symptoms revealing short circuits included the wearing off of therapeutic effect, apraxia of eyelid opening, or dysarthria in 6 patients with Parkinson disease (PD), and dystonia deterioration in 1 patient with generalized dystonia. All DBS leads with short circuits had been anchored to the cranium using titanium miniplates. Altering electrode settings resulted in clinical improvement in the 2 PD cases in which patients had specific symptoms of short circuits (2.5%) but not in the other 4 cases. The patient with dystonia underwent repositioning and replacement of a lead because the previous lead was located too anteriorly, but did not experience symptom improvement.

Conclusions. In contrast to the sudden loss of clinical efficacy of DBS caused by an open circuit, short circuits may arise due to a gradual decrease in impedance, causing the insidious development of neurological symptoms via limited or extended potential fields as well as shortened battery longevity. The incidence of short circuits in DBS may be higher than previously thought, especially in cases in which DBS leads are anchored with miniplates. The circuit impedance of DBS should be routinely checked, even after a long history of DBS therapy, especially in cases of miniplate anchoring.

Intraoperative magnetic resonance imaging findings during deep brain stimulation surgery

J Neurosurg 115:852–857, 2011.DOI: 10.3171/2011.5.JNS101457

Deep brain stimulation (DBS) is an established neurosurgical technique used to treat a variety of neurological disorders, including Parkinson disease, essential tremor, dystonia, epilepsy, depression, and obsessive-compulsive disorder. This study reports on the use of intraoperative MR imaging during DBS surgery to evaluate acute hemorrhage, intracranial air, brain shift, and accuracy of lead placement.

Methods. During a 46-month period, 143 patients underwent 152 DBS surgeries including 289 lead placements utilizing intraoperative 1.5-T MR imaging. Imaging was supervised by an MR imaging physicist to maintain the specific absorption rate below the required level of 0.1 W/kg and always included T1 magnetization-prepared rapid gradient echo and T2* gradient echo sequences with selected use of T2 fluid attenuated inversion recovery (FLAIR) and T2 fast spin echo (FSE). Retrospective review of the intraoperative MR imaging examinations was performed to quantify the amount of hemorrhage and the amount of air introduced during the DBS surgery.

Results. Intraoperative MR imaging revealed 5 subdural hematomas, 3 subarachnoid hemorrhages, and 1 intraparenchymal hemorrhage in 9 of the 143 patients. Only 1 patient experiencing a subarachnoid hemorrhage developed clinically apparent symptoms, which included transient severe headache and mild confusion. Brain shift due to intracranial air was identified in 144 separate instances.

Conclusions. Intraoperative MR imaging can be safely performed and may assist in demonstrating acute changes involving intracranial hemorrhage and air during DBS surgery. These findings are rarely clinically significant and typically resolve prior to follow-up imaging. Selective use of T2 FLAIR and T2 FSE imaging can confirm the presence of hemorrhage or air and preclude the need for CT examinations

Frameless deep brain stimulation using intraoperative O-arm technology

J Neurosurg 115:301–309, 2011. DOI: 10.3171/2011.3.JNS101642

Correct lead location in the desired target has been proven to be a strong influential factor for good clinical outcome in deep brain stimulation (DBS) surgery. Commonly, a surgeon’s first reliable assessment of such location is made on postoperative imaging. While intraoperative CT (iCT) and intraoperative MR imaging have been previously described, the authors present a series of frameless DBS procedures using O-arm iCT.

Methods. Twelve consecutive patients with 15 leads underwent frameless DBS placement using electrophysiological testing and O-arm iCT. Initial target coordinates were made using standard indirect and direct assessment. Microelectrode recording (MER) with kinesthetic responses was performed, followed by microstimulation to evaluate the side-effect profile. Intraoperative 3D CT acquisitions obtained between each MER pass and after final lead placement were fused with the preoperative MR image to verify intended MER movements around the target area and to identify the final lead location. Tip coordinates from the initial plan, final intended target, and actual lead location on iCT were later compared with the lead location on postoperative MR imaging, and euclidean distances were calculated. The amount of radiation exposure during each procedure was calculated and compared with the estimated radiation exposure if iCT was not performed.

Results. The mean euclidean distances between the coordinates for the initial plan, final intended target, and actual lead on iCT compared with the lead coordinates on postoperative MR imaging were 3.04 ± 1.45 mm (p = 0.0001), 2.62 ± 1.50 mm (p = 0.0001), and 1.52 ± 1.78 mm (p = 0.0052), respectively. The authors obtained good merging error during image fusion, and postoperative brain shift was minimal. The actual radiation exposure from iCT was invariably less than estimates of exposure using standard lateral fluoroscopy and anteroposterior radiographs (p < 0.0001).

Conclusions. O-arm iCT may be useful in frameless DBS surgery to approximate microelectrode or lead locations intraoperatively. Intraoperative CT, however, may not replace fundamental DBS surgical techniques such as electrophysiological testing in movement disorder surgery. Despite the lack of evidence for brain shift from the procedure, iCT-measured coordinates were statistically different from those obtained postoperatively, probably indicating image merging inaccuracy and the difficulties in accurately denoting lead location. Therefore, electrophysiological testing may truly be the only means of precisely knowing the location in 3D space intraoperatively. While iCT may provide clues to electrode or lead location during the procedure, its true utility may be in DBS procedures targeting areas where electrophysiology is less useful. The use of iCT appears to reduce radiation exposure compared with the authors’ traditional frameless technique.