Frameless deep brain stimulation using intraoperative O-arm technology

J Neurosurg 115:301–309, 2011. DOI: 10.3171/2011.3.JNS101642

Correct lead location in the desired target has been proven to be a strong influential factor for good clinical outcome in deep brain stimulation (DBS) surgery. Commonly, a surgeon’s first reliable assessment of such location is made on postoperative imaging. While intraoperative CT (iCT) and intraoperative MR imaging have been previously described, the authors present a series of frameless DBS procedures using O-arm iCT.

Methods. Twelve consecutive patients with 15 leads underwent frameless DBS placement using electrophysiological testing and O-arm iCT. Initial target coordinates were made using standard indirect and direct assessment. Microelectrode recording (MER) with kinesthetic responses was performed, followed by microstimulation to evaluate the side-effect profile. Intraoperative 3D CT acquisitions obtained between each MER pass and after final lead placement were fused with the preoperative MR image to verify intended MER movements around the target area and to identify the final lead location. Tip coordinates from the initial plan, final intended target, and actual lead location on iCT were later compared with the lead location on postoperative MR imaging, and euclidean distances were calculated. The amount of radiation exposure during each procedure was calculated and compared with the estimated radiation exposure if iCT was not performed.

Results. The mean euclidean distances between the coordinates for the initial plan, final intended target, and actual lead on iCT compared with the lead coordinates on postoperative MR imaging were 3.04 ± 1.45 mm (p = 0.0001), 2.62 ± 1.50 mm (p = 0.0001), and 1.52 ± 1.78 mm (p = 0.0052), respectively. The authors obtained good merging error during image fusion, and postoperative brain shift was minimal. The actual radiation exposure from iCT was invariably less than estimates of exposure using standard lateral fluoroscopy and anteroposterior radiographs (p < 0.0001).

Conclusions. O-arm iCT may be useful in frameless DBS surgery to approximate microelectrode or lead locations intraoperatively. Intraoperative CT, however, may not replace fundamental DBS surgical techniques such as electrophysiological testing in movement disorder surgery. Despite the lack of evidence for brain shift from the procedure, iCT-measured coordinates were statistically different from those obtained postoperatively, probably indicating image merging inaccuracy and the difficulties in accurately denoting lead location. Therefore, electrophysiological testing may truly be the only means of precisely knowing the location in 3D space intraoperatively. While iCT may provide clues to electrode or lead location during the procedure, its true utility may be in DBS procedures targeting areas where electrophysiology is less useful. The use of iCT appears to reduce radiation exposure compared with the authors’ traditional frameless technique.

Intraoperative Computed Tomography for Deep Brain Stimulation Surgery: Technique and Accuracy Assessment

Neurosurgery 68[ONS Suppl 1]:ons114–ons124, 2011. DOI: 10.1227/NEU.0b013e31820781bc

The efficacy of deep brain stimulation (DBS) is highly dependent on the accuracy of lead placement.

OBJECTIVE: To describe the use of intraoperative computed tomography (iCT) to confirm lead location before surgical closure and to study the accuracy of this technique.

METHODS: Fifteen patients underwent awake microelectrode-guided DBS surgery in a stereotactic frame. A portable iCT scanner (Medtronic O-arm) was positioned around the patient’s head throughout the procedure and was used to confirm lead location before fixation of the lead to the skull. Images were computationally fused with preoperative magnetic resonance imaging (MRI), and lead tip coordinates with respect to the midpoint of the anterior commissure-posterior commissure line were measured. Tip coordinates were compared with those obtained from postoperative MRI.

RESULTS: iCT was integrated into standard frame-based microelectrode-guided DBS surgery with a minimal increase in surgical time or complexity. Technically adequate 2-dimensional and 3-dimensional images were obtained in all cases. Head positioning and fixation techniques that allow unobstructed imaging are described. Lead tip measurements on iCT fused with preoperative MRI were statistically indistinguishable from those obtained with postoperative MRI.

CONCLUSION: iCT can be easily incorporated into standard DBS surgery, replaces the need for C-arm fluoroscopy, and provides accurate intraoperative 3-dimensional confirmation of electrode tip locations relative to preoperative images and surgical plans. iCT fused to preoperative MRI may obviate the need for routine postoperative MRI in DBS surgery. Technical nuances that must be mastered for the efficient use of iCT during DBS implantation are described.