Shunt Failure—The First 30 Days

Neurosurgery 87:123–129, 2020

Incontrovertible predictors of shunt malfunction remain elusive.

OBJECTIVE: To determine predictors of shunt failure within 30 d of index surgery.

METHODS: This was a single-center retrospective cohort study from January 2010 through November 2016. Using a ventricular shunt surgery research database, clinical and procedural variables were procured. An “index surgery” was defined as implantation of a new shunt or revision or augmentation of an existing shunt system. The primary outcome was shunt failure of any kind within the first 30 days of index surgery. Bivariate models were created, followed by a final multivariable logistic regression model using a backward-forward selection procedure.

RESULTS: Our dataset contained 655 unique patients with a total of 1206 operations. The median age for the cohort at the time of first shunt surgery was 4.6 yr (range, 0-28; first and third quartile, .37 and 11.8, respectively). The 30-day failure rates were 12.4% when analyzing the first-index operation only (81/655), and 15.7% when analyzing all-index operations (189/1206). Small or slit ventricles at the time of index surgery and prior ventricular shunt operations were found to be significant covariates in both the “first-index” (P < .01 and P=.05, respectively) and “all-index”(P=.02 and P<.01, respectively) multivariable models. Intraventricular hemorrhage at the time of index surgery was an additional predictor in the all-index model (P = .01).

CONCLUSION: This study demonstrates that only 3 variables are predictive of 30-day shunt failure when following established variable selection procedures, 2 of which are potentially under direct control of the surgeon.

More malfunctioning Medos Hakim programmable valves: cause for concern?

J Neurosurg 115:1047–1052, 2011. DOI: 10.3171/2011.5.JNS101396

In recent years, the authors have noticed a growing number of programmable valve defects at their institution. Therefore, they conducted this study to evaluate the increased incidence of malfunctioning valves.

Methods. They investigated all revisions that had been performed at their institution between 1994 and 2010 for dislodgement of the stator of a standard Medos Hakim programmable valve with a prechamber.

Results. Fifteen valves were removed because of dislodged stators. The valves had been implanted between May 16, 1993, and December 27, 2002, and were explanted between February 19, 2006, and January 22, 2010. Thus, the valves had been in place for a mean period of 11 years (median 11 years, range 7–14 years). The percentage of dislodged stators was almost 3% (15 of 546 valves). Particularly noteworthy is that all malfunctioning valves were found in children who had been younger than 1 year of age at the time of implantation.

Conclusions. Medos Hakim programmable valve malfunctions are rare events but should receive careful attention. When the pressure setting cannot be adjusted, a malfunction should always be suspected and radiographic imaging should be performed to assess the valve. Stator dislodgement is the most serious form of valve adjustment failure.