Navigated bedside implantation of external ventricular drains with mobile health guidance

Acta Neurochirurgica (2024) 166:76

External ventricular drain (EVD) implantation is one of the fundamental procedures of emergency neurosurgery usually performed freehand at bedside or in the operating room using anatomical landmarks. However, this technique is frequently associated with malpositioning leading to complications or dysfunction. Here, we describe a novel navigated bedside EVD insertion technique, which is evaluated in a clinical case series with the aim of safety, accuracy, and efficiency in neurosurgical emergency settings.

Methods From 2021 to 2022, a mobile health–assisted navigation instrument (Thomale Guide, Christoph Miethke, Potsdam, Germany) was used alongside a battery-powered single-use drill (Phasor Health, Houston, USA) for bedside EVD placement in representative neurosurgical pathologies in emergency situations requiring ventricular cerebrospinal fluid (CSF) relief and intracranial pressure (ICP) monitoring.

Results In all 12 patients (8 female and 4 male), navigated bedside EVDs were placed around the foramen of Monro at the first ventriculostomy attempt. The most frequent indication was aneurysmal subarachnoid hemorrhage. Mean operating time was 25.8 ± 15.0 min. None of the EVDs had to be revised due to malpositioning or dysfunction. Two EVDs were converted into a ventriculoperitoneal shunt. Drainage volume was 41.3 ± 37.1 ml per day in mean. Mean length of stay of an EVD was 6.25 ± 2.8 days. Complications included one postoperative subdural hematoma and cerebrospinal fluid infection, respectively.

Conclusion Combining a mobile health–assisted navigation instrument with a battery-powered drill and an appropriate ventricular catheter may enable and enhance safety, accuracy, and efficiency in bedside EVD implantation in various pathologies of emergency neurosurgery without adding relevant efforts.

GAVCA Study: Randomized, Multicenter Trial to Evaluate the Quality of Ventricular Catheter Placement with a Mobile Health Assisted Guidance Technique

Neurosurgery 83:252–262, 2018

Freehand ventricular catheter placementmay represent limited accuracy for the surgeon’s intent to achieve primary optimal catheter position.

OBJECTIVE: To investigate the accuracy of a ventricular catheter guide assisted by a simple mobile health application (mhealth app) in a multicenter, randomized, controlled, simple blinded study (GAVCA study).

METHODS: In total, 139 eligible patients were enrolled in 9 centers. Catheter placement was evaluated by 3 different components: number of ventricular cannulation attempts, a grading scale, and the anatomical position of the catheter tip. The primary endpoint was the rate of primary cannulation of grade I catheter position in the ipsilateral ventricle. The secondary endpoints were rate of intraventricular position of the catheter’s perforations, early ventricular catheter failure, and complications.

RESULTS: The primary endpoint was reached in 70% of the guided group vs 56.5% (freehand group; odds ratio 1.79, 95% confidence interval 0.89-3.61). The primary successful puncture rate was 100% vs 91.3% (P= .012). Catheter perforations were located completely inside the ventricle in 81.4% (guided group) and 65.2% (freehand group; odds ratio 2.34, 95% confidence interval 1.07-5.1). No differences occurred in early ventricular catheter failure, complication rate, duration of surgery, or hospital stay.

CONCLUSION: The guided ventricular catheter application proved to be a safe and simple method. The primary endpoint revealed a nonsignificant improvement of optimal catheter placement among the groups. Long-term follow-up is necessary in order to evaluate differences in catheter survival among shunted patients.