Ability of electromyographic monitoring to determine the presence of malpositioned pedicle screws in the lumbosacral spine: analysis of 2450 consecutively placed screws

J Neurosurg Spine 15:130–135, 2011.DOI: 10.3171/2011.3.SPINE101

Pedicle screws provide efficient stabilization along all 3 columns of the spine, but they can be technically demanding to place, with malposition rates ranging from 5% to 10%. Intraoperative electromyographic (EMG) monitoring has the capacity to objectively identify a screw breaching the medial pedicle cortex that is in proximity to a nerve root. The purpose of this study is to describe and evaluate the authors’ 7-year institutional experience with intraoperative EMG monitoring during placement of lumbar pedicle screws and to determine the clinical utility of intraoperative EMG monitoring.

Methods. The authors retrospectively studied 2450 consecutive lumbar pedicle screws placed in 418 patients from June 2002 through June 2009. All screws were inserted using a free-hand technique and anatomical landmarks, stimulated at 10.0 mA, and evaluated with CT scanning within 48 hours postoperatively. Medial pedicle screw breach was defined as having greater than 25% of the screw diameter extend outside of the pedicle, as confirmed on CT scanning or intraoperatively by a positive EMG response indicating a medial breach. The sensitivity and specificity of intraoperative EMG monitoring in detecting the presence of a medial screw breach was evaluated based on the following definitions: 1) true positive (a positive response to EMG stimulation confirmed as a breach intraoperatively or on postoperative CT scans); 2) false positive (positive response to EMG stimulation confirmed as a correctly positioned screw on postoperative CT scans); 3) true negative (no response to EMG stimulation confirmed as a correctly positioned screw on postoperative CT scans); or 4) false negative (no response to EMG stimulation but confirmed as a breach on postoperative CT scans).

Results. One hundred fifteen pedicle screws (4.7%) showed positive stimulation during intraoperative EMG monitoring. At stimulation thresholds less than 5.0, 5.0–8.0, and > 8.0 mA, the specificity of a positive response was 99.9%, 97.9%, and 95.9%, respectively. The sensitivity of a positive response at these thresholds was only 43.4%, 69.6%, and 69.6%, respectively. At a threshold less than 5.0 mA, 91% of screws with a positive EMG response were confirmed as true medial breaches. However, at thresholds of 5.0–8.0 mA or greater than 8.0 mA, a positive EMG response was associated with 89% and 100% false positives (no breaches), respectively.

Conclusions. When using intraoperative EMG monitoring, a positive response at screw stimulation thresholds less than 5.0 mA was highly specific for a medial pedicle screw breach but was poorly sensitive. A positive response to stimulation thresholds greater 5.0 mA was associated with a very high rate of false positives. The authors’ experience suggests that pedicle screws showing positive stimulation below 5.0 mA warrants intraoperative investigation for malpositioning while responses at higher thresholds are less reliable at accurately representing a medial breach

Minimally invasive percutaneous transpedicular screw fixation: increased accuracy and reduced radiation exposure by means of a novel electromagnetic navigation system

Acta Neurochir (2011) 153:589–596. DOI 10.1007/s00701-010-0882-4

Minimally invasive percutaneous pedicle screw instrumentation methods may increase the need for intraoperative fluoroscopy, resulting in excessive radiation exposure for the patient, surgeon, and support staff. Electromagnetic field (EMF)-based navigation may aid more accurate placement of percutaneous pedicle screws while reducing fluoroscopic exposure. We compared the accuracy, time of insertion, and radiation exposure of EMF with traditional fluoroscopic percutaneous pedicle screw placement.

Methods Minimally invasive pedicle screw placement in T8 to S1 pedicles of eight fresh-frozen human cadaveric torsos was guided with EMF or standard fluoroscopy. Setup, insertion, and fluoroscopic times and radiation exposure and accuracy (measured with post-procedural computed tomography) were analyzed in each group.

Results Sixty-two pedicle screws were placed under fluoroscopic guidance and 60 under EMF guidance. Ideal trajectories were achieved more frequently with EMF over all segments (62.7% vs. 40%; p=0.01). Greatest EMF accuracy was achieved in the lumbar spine, with significant improvements in both ideal trajectory and reduction of pedicle breaches over fluoroscopically guided placement (64.9% vs. 40%, p=0.03, and 16.2% vs. 42.5%, p=0.01, respectively). Fluoroscopy time was reduced 77% with the use of EMF (22 s vs. 5 s per level; p<0.0001) over all spinal segments. Radiation exposure at the hand and body was reduced 60% (p=0.058) and 32% (p=0.073), respectively. Time for insertion did not vary between the two techniques.

Conclusions Minimally invasive pedicle screw placement with the aid of EMF image guidance reduces fluoroscopy time and increases placement accuracy when compared with traditional fluoroscopic guidance while adding no additional time to the procedure.

Accuracy of Free-Hand Pedicle Screws in the Thoracic and Lumbar Spine: Analysis of 6816 Consecutive Screws

Neurosurgery 68:170–178, 2011 DOI: 10.1227/NEU.0b013e3181fdfaf4

Pedicle screws are used to stabilize all 3 columns of the spine, but can be technically demanding to place. Although intraoperative fluoroscopy and stereotactic- guided techniques slightly increase placement accuracy, they are also associated with increased radiation exposure to patient and surgeon as well as increased operative time.

OBJECTIVE: To describe and critically evaluate our 7-year institutional experience with placement of pedicle screws in the thoracic and lumbar spine using a free-hand technique.

METHODS: We retrospectively reviewed records of all patients undergoing free-hand pedicle screw placement without fluoroscopy in the thoracic or lumbar spine between June 2002 and June 2009. Incidence and extent of cortical breach by misplaced pedicle screw was determined by review of postoperative computed tomography scans. We defined breach as more than 25% of the screw diameter residing outside of the pedicle or vertebral body cortex.

RESULTS: A total of 964 patients received 6816 free-hand placed pedicle screws in the thoracic or lumbar spine. Indications for hardware placement were degenerative/deformity disease (51.2%), spondylolisthesis (23.7%), tumor (22.7%), trauma (11.3%), infection (7.6%), and congenital (0.9%). A total of 115 screws (1.7%) were identified as breaching the pedicle in 87 patients (9.0%). Breach occurred more frequently in the thoracic than the lumbar spine (2.5% and 0.9%, respectively; P , .0001) and was more often lateral (61.3%) than medial (32.8%) or superior (2.5%). T4 (4.1%) and T6 (4.0%) experienced the highest breach rate, whereas L5 and S1 had the lowest breach rate. Eight patients (0.8%) underwent revision surgery to correct malpositioned screws.

CONCLUSION: Free-hand pedicle screw placement based on external anatomy alone can be performed with acceptable safety and accuracy and allows avoidance of radiation exposure encountered in fluoroscopic techniques. Image-guided assistance may be most valuable when placing screws between T4 and T6, where breach rates are highest

A retrospective analysis of pedicle screws in contact with the great vessels

J Neurosurg Spine 13:403–406, 2010.DOI: 10.3171/2010.3.SPINE09657

Pedicle screws placed in the thoracic, lumbar, and sacral spine occasionally come in contact with the aorta, vena cava, or iliac vessels. When such screws are seen on postoperative imaging in an asymptomatic patient, the surgeon must decide whether it is riskier to revise the screw or to observe it. The authors hypothesized that the incidence of screw placement causing perioperative vessel injury is low and, further, that screws placed in contact with major vessels do not always result in vessel injury.

Methods. A retrospective review of the operative records of 182 consecutive patients undergoing thoracic, lumbar, and lumbosacral pedicle screw fusion was performed to determine the frequency of intraoperative vessel injury. Postoperative imaging for 107 patients was available to determine the incidence of screws in contact with major vessels. Charts were examined to determine if any adverse sequelae had resulted from malpositioned screws. Patient outcomes were documented.

Results. There were no intraoperative vessel injuries or deaths in 182 consecutive operations. One hundred seven patients with available postoperative films had 680 pedicle screws placed between T-3 and the sacrum during 115 operations. No patient had arterial screw penetration or deformation on postoperative imaging. Thirty-three of the 680 inserted screws were in contact with a major vessel on routine postoperative imaging. The contacted vessels included the aorta (4 cases), the iliac artery (7 cases), and the iliac veins (22 cases). Patients were followed up until death or November 2009, for a mean follow-up of 44 months (median 44 months, range 5–109 months). None of the patients with vessel contact was noted to suffer symptoms or sequelae as a result of vessel contact. Radiographic follow-up as long as 50 months after surgery revealed no detectable vessel abnormality at the contacted site.

Conclusions. Placing pedicle screws in contact with major vessels is a known risk of spinal surgery. The risk of repositioning a screw in contact with a major vessel but causing no symptoms must be weighed against the relative risk of leaving it in place.

An alternate method for placement of C-1 screws

J Neurosurg: Spine / Volume 12 / April 2010. DOI: 10.3171/2009.10.SPINE08541

Several techniques for the surgical stabilization of the atlas and the axis have been described. Placement of C-1 lateral mass screws is one of the latest technical advances, and has gained popularity due to its efficacy and biomechanical advantages. However, the technique for placement of C-1 lateral mass screws, as first described by Harms, can cause excessive bleeding or irritation of the C-2 nerve. An alternative technique is available for the placement of C-1 lateral mass screws that completely avoids the C-2 nerve/ganglion and its associated venous plexus. This new technique mitigates some of the risk associated with the Harms techniques and eliminates the need to use specialized screws (that is, smooth shanks).

Methods: Twenty-six patients underwent atlantoaxial or occipitocervical fusions incorporating the alternative technique of C-1 screw placement. Three surgeons at 3 different institutions performed the surgeries. Standard lateral fluoroscopy and fully threaded polyaxial screws were used in each case.

Results: Forty-nine screws were placed in C-1 lateral masses by using the new technique. Solid arthrodesis was achieved in all cases, with a mean follow-up period of 30 months. There were no cases of CSF leakage, new neurological deficit, injury to the C-2 ganglion, vertebral artery injury, or hardware failures.

Conclusions: The technique is a safe and effective way to fixate C-1 while avoiding the C-2 nerve/ganglion and venous plexus. The results indicate that excellent clinical and radiographic outcomes can be achieved with this new technique.

Accuracy of pedicle screw placement in the lumbosacral spine using conventional technique: CT postoperative assessment in 102 consecutive patients

Journal of Neurosurgery: Spine. March 2010.DOI: 10.3171/2009.9.SPINE09261

The goal of this study was to determine the incidence of screw misplacement and complications in a group of 102 patients who underwent transpedicle screw fixation in the lumbosacral spine with conventional open technique and intraoperative fluoroscopy. The results are compared with published data.

Methods: Cases involving 102 consecutive patients (424 inserted screws) were reviewed. Surgery was performed in all cases by the same surgeon’s team, using the same implant, and all results were assessed by means of a specific CT protocol. The screw position was assessed by the authors and an independent observer. Screw position was classified as correct when the screw was completely surrounded by the pedicle cortex, as “cortical encroachment” (questionable violation) if the pedicle cortex could not be visualized, and as “frank penetration” when the screw was outside the pedicular boundaries. Frank penetration was further subdivided as minor (when the edge of the screw thread was up to 2.0 mm outside the pedicle cortex), moderate (2.1–4 mm), and severe (> 4 mm). The incidence of intra- and postoperative complications not related to screw position as well as hardware failures were also registered, with a minimum follow-up duration of 8 months.

Results: The rate of frank pedicle screw misplacement was 5%. The rate of minimal or questionable pedicle wall violation was 2.8%. Among the frank misplacements, 6 were classified as minor, 12 as moderate, and 3 as severe penetration. Two patients (2%) had radicular pain and neurological deficits (inferomedial and inferolateral minor misplacement at L-4 and L-5, respectively), and 5 patients (4.9%) complained only of radicular pain. At the follow-up examination all patients had completely recovered their neurological function and radicular pain was resolved in all cases. The complications not related to screw malposition were 2 pedicle fractures (2% of patients), 1 nerve root injury (1%), and 1 dural laceration (1%). Five patients (4.8%) had postoperative anemia and required transfusions. Superficial or deep wound infection was noted in 3 patients (2.9%). Late hardware failure occurred in 2 patients (2%). One patient developed adjacent segmental instability and required additional surgery to extend the fusion.

Conclusions: Our rates of screw misplacement and complications compare favorably with the lowest rates of the series in which conventional technique was used and are close to the rates reported for image-guided methods. The risk of malpositioning may be reduced with careful preoperative surgical planning, accurate knowledge of the spinal anatomy, surgical experience, and correct indication for conventional surgery. The conventional technique still remains a practical, safe, and effective surgical method for lumbosacral fixation.