Robot-Assisted Minimally Invasive Asleep Single-Stage Deep Brain Stimulation Surgery

Operative Neurosurgery 26:363–371, 2024

Robotic assistance has garnered increased use in neurosurgery. Recently, this has expanded to include deep brain stimulation (DBS). Several studies have reported increased accuracy and improved efficiency with robotic assistance, but these are limited to individual robotic platforms with smaller sample sizes or are broader studies on robotics not specific to DBS. Our objectives are to report our technique for robot-assisted, minimally invasive, asleep, single-stage DBS surgery and to perform a meta-analysis comparing techniques from previous studies.

METHODS: We performed a single-center retrospective review of DBS procedures using a floor-mounted robot with a frameless transient fiducial array registration. We compiled accuracy data (radial entry error, radial target error, and 3dimensional target error) and efficiency data (operative time, setup time, and total procedure time). We then performed a meta-analysis of previous studies and compared these metrics.

RESULTS: We analyzed 315 electrodes implanted in 160 patients. The mean radial target error was 0.9 ± 0.5 mm, mean target 3-dimensional error was 1.3 ± 0.7 mm, and mean radial entry error was 1.1 ± 0.8 mm. The mean procedure time (including pulse generator placement) was 182.4 ± 47.8 minutes, and the mean setup time was 132.9 ± 32.0 minutes. The overall complication rate was 8.8% (2.5% hemorrhagic/ischemic, 2.5% infectious, and 0.6% revision). Our meta-analysis showed increased accuracy with floor-mounted over skull-mounted robotic platforms and with fiducial-based registrations over optical registrations.

CONCLUSION: Our technique for robot-assisted, minimally invasive, asleep, single-stage DBS surgery is safe, accurate, and efficient. Our data, combined with a meta-analysis of previous studies, demonstrate that robotic assistance can provide similar or increased accuracy and improved efficiency compared with traditional frame-based techniques. Our analysis also suggests that floor-mounted robots and fiducial-based registration methods may be more accurate.

 

A New Noninvasive Frameless Registration System for Stereotactic Cranial Biopsy

 Operative Neurosurgery 24:64–67, 2023

Although frame-based stereotactic biopsy is still considered the gold standard for brain biopsies, frameless robot-assisted stereotactic systems are now able to provide an equal level of safety and accuracy. However, both systems suffer from a lack of efficiency of the operative workflow.

OBJECTIVE: To describe the technique of a new frameless and noninvasive registration tool Neurolocate (Renishaw). This tool, combined with an intraoperative cone-beam computed tomography imaging system like O-ARM (Medtronic), might facilitate the achievement and workflow of robot-assisted stereotactic intracranial biopsies.

METHODS: Neurolocate is a 3-dimensional fiducial tool fixed directly on the Neuromate (Renishaw) robot arm. It consists of 5 radio-opaque spherical fiducials, whose geometry is constant. This tool made it possible to carry out the coregistration then the biopsy in the same operating time, following a five-step procedure described here. We retrospectively extracted selected preliminary results from our initial experience.

RESULTS: Over 1 year, 23 consecutive adult patients were biopsied with Neurolocate in our center. The mean overall operative time, from patient’s installation to skin closure, was 97 minutes ± 27 (SD). The entire procedure took place in a single location unit (operating room), which facilitated workflow and surgical planning. No invasive gesture was performed outside of the operating time.

CONCLUSION: Neurolocate is a new frameless and noninvasive registration tool that could improve workflow and flexibility for operating room management and surgical planning. It may also increase the comfort of patients undergoing robot-assisted intracranial stereotactic biopsies. The accuracy and safety profile should be addressed in specific studies.

 

A novel miniature robotic device for frameless implantation of depth electrodes in refractory epilepsy

J Neurosurg 126:1622–1628, 2017

The authors’ group recently published a novel technique for a navigation-guided frameless stereotactic approach for the placement of depth electrodes in epilepsy patients. To improve the accuracy of the trajectory and enhance the procedural workflow, the authors implemented the iSys1 miniature robotic device in the present study into this routine.

METHODS As a first step, a preclinical phantom study was performed using a human skull model, and the accuracy and timing between 5 electrodes implanted with the manual technique and 5 with the aid of the robot were compared. After this phantom study showed an increased accuracy with robot-assisted electrode placement and confirmed the robot’s ability to maintain stability despite the rotational forces and the leverage effect from drilling and screwing, patients were enrolled and analyzed for robot-assisted depth electrode placement at the authors’ institution from January 2014 to December 2015. All procedures were performed with the S7 Surgical Navigation System with Synergy Cranial software and the iSys1 miniature robotic device.

RESULTS Ninety-three electrodes were implanted in 16 patients (median age 33 years, range 3–55 years; 9 females, 7 males). The authors saw a significant increase in accuracy compared with their manual technique, with a median deviation from the planned entry and target points of 1.3 mm (range 0.1–3.4 mm) and 1.5 mm (range 0.3–6.7 mm), respectively. For the last 5 patients (31 electrodes) of this series the authors modified their technique in placing a guide for implantation of depth electrodes (GIDE) on the bone and saw a significant further increase in the accuracy at the entry point to 1.18 ± 0.5 mm (mean ± SD) compared with 1.54 ± 0.8 mm for the first 11 patients (p = 0.021). The median length of the trajectories was 45.4 mm (range 19–102.6 mm). The mean duration of depth electrode placement from the start of trajectory alignment to fixation of the electrode was 15.7 minutes (range 8.5–26.6 minutes), which was significantly faster than with the manual technique. In 12 patients, depth electrode placement was combined with subdural electrode placement. The procedure was well tolerated in all patients. The authors did not encounter any case of hemorrhage or neurological deficit related to the electrode placement. In 1 patient with a psoriasis vulgaris, a superficial wound infection was encountered. Adequate physiological recordings were obtained from all electrodes. No additional electrodes had to be implanted because of misplacement.

CONCLUSIONS The iSys1 robotic device is a versatile and easy to use tool for frameless implantation of depth electrodes for the treatment of epilepsy. It increased the accuracy of the authors’ manual technique by 60% at the entry point and over 30% at the target. It further enhanced and expedited the authors’ procedural workflow.

Minimally invasive transforaminal lumbar interbody fusion with the ROSA Spine robot and intraoperative flat-panel CT guidance

Minimally invasive transforaminal lumbar interbody fusion with the ROSATM Spine robot and intraoperative flat-panel CT guidance

Acta Neurochir (2016) 158:1125–1128

Circumferential arthrodesis is commonly used to treat degenerative lumbar diseases. Minimally invasive techniques may enable faster recovery and reduce the incidence of postoperative infections.

Methods: We report on the surgical technique of a transforaminal lumbar interbody fusion (TLIF) procedure performed with the assistance of a new robotic device (ROSATM Spine) and intraoperative flat-panel CT guidance.

Conclusions: The combined use of this new robotic device and intraoperative CT enables accurate and safe arthrodesis in the treatment of degenerative lumbar disc diseases.

Robotics and the spine: a review of current and ongoing applications

Virtual_Reality_and_Robotics_in

Neurosurg Focus 36 (3):E10, 2014

Robotics in the operating room has shown great use and versatility in multiple surgical fields. Robotassisted spine surgery has gained significant favor over its relatively short existence, due to its intuitive promise of higher surgical accuracy and better outcomes with fewer complications. Here, the authors analyze the existing literature on this growing technology in the era of minimally invasive spine surgery.

Methods. In an attempt to provide the most recent, up-to-date review of the current literature on robotic spine surgery, a search of the existing literature was conducted to obtain all relevant studies on robotics as it relates to its application in spine surgery and other interventions.

Results. In all, 45 articles were included in the analysis. The authors discuss the current status of this technology and its potential in multiple arenas of spinal interventions, mainly spine surgery and spine biomechanics testing.

Conclusions. There are numerous potential advantages and limitations to robotic spine surgery, as suggested in published case reports and in retrospective and prospective studies. Randomized controlled trials are few in number and show conflicting results regarding accuracy. The present limitations may be surmountable with future technological improvements, greater surgeon experience, reduced cost, improved operating room dynamics, and more training of surgical team members. Given the promise of robotics for improvements in spine surgery and spine biomechanics testing, more studies are needed to further explore the applicability of this technology in the spinal operating room. Due to the significant cost of the robotic equipment, studies are needed to substantiate that the increased equipment costs will result in significant benefits that will justify the expense.