Rod fractures in thoracolumbar fusions to the sacrum/pelvis for adult symptomatic lumbar scoliosis

J Neurosurg Spine 38:217–229, 2023

Previous reports of rod fracture (RF) in adult spinal deformity are limited by heterogeneous cohorts, low follow-up rates, and relatively short follow-up durations. Since the majority of RFs present > 2 years after surgery, true occurrence and revision rates remain unclear. The objectives of this study were to better understand the risk factors for RF and assess its occurrence and revision rates following primary thoracolumbar fusions to the sacrum/pelvis for adult symptomatic lumbar scoliosis (ASLS) in a prospective series with long-term follow-up.

METHODS Patient records were obtained from the Adult Symptomatic Lumbar Scoliosis–1 (ASLS-1) database, an NIH-sponsored multicenter, prospective study. Inclusion criteria were as follows: patients aged 40–80 years undergoing primary surgeries for ASLS (Cobb angle ≥ 30° and Oswestry Disability Index ≥ 20 or Scoliosis Research Society-22r ≤ 4.0 in pain, function, and/or self-image) with instrumented fusion of ≥ 7 levels that included the sacrum/pelvis. Patients with and without RF were compared to assess risk factors for RF and revision surgery.

RESULTS Inclusion criteria were met by 160 patients (median age 62 years, IQR 55.7–67.9 years). At a median followup of 5.1 years (IQR 3.8–6.6 years), there were 92 RFs in 62 patients (38.8%). The median time to RF was 3.0 years (IQR 1.9–4.54 years), and 73% occurred > 2 years following surgery. Based on Kaplan-Meier analyses, estimated RF rates at 2, 4, 5, and 8 years after surgery were 11%, 24%, 35%, and 49%, respectively. Baseline radiographic, clinical, and demographic characteristics were similar between patients with and without RF. In Cox regression models, greater postoperative pelvic tilt (HR 1.895, 95% CI 1.196–3.002, p = 0.0065) and greater estimated blood loss (HR 1.02, 95% CI 1.005–1.036, p = 0.0088) were associated with increased risk of RF. Thirty-eight patients (61% of all RFs) underwent revision surgery. Bilateral RF was predictive of revision surgery (HR 3.52, 95% CI 1.8–6.9, p = 0.0002), while patients with unilateral nondisplaced RFs were less likely to require revision (HR 0.39, 95% CI 0.18–0.84, p = 0.016).

CONCLUSIONS This study provides what is to the authors’ knowledge the highest-quality data to date on RF rates following ASLS surgery. At a median follow-up of 5.1 years, 38.8% of patients had at least one RF. Estimated RF rates at 2, 4, 5, and 8 years after surgery were 11%, 24%, 35%, and 49%, respectively. Greater estimated blood loss and postoperative pelvic tilt were significant risk factors for RF. These findings emphasize the importance of long-term follow-up to realize the true prevalence and cumulative incidence of RF.

Posterior fusion for fragility type 2 odontoid fractures

J Neurosurg Spine 35:644–650, 2021

The purpose of this study was to evaluate the outcomes of elderly patients with type 2 odontoid fractures treated with an instrumented posterior fusion.

METHODS Ninety-three consecutive patients older than 65 years of age in whom a type 2 odontoid fracture had been treated with a variety of C1–2 posterior screw fixation techniques were retrospectively reviewed.

RESULTS The average age was 78 years (range 65–95 years). Thirty-seven patients had an additional fracture, 30 of which involved C1. Three patients had cervical spinal cord dysfunction due to their injury. All patients had comorbidities. The average total hospitalization was 9.6 days (range 2–37 days). There were 3 deaths and 19 major complications, the most common of which was pneumonia. No patient suffered a vertebral artery injury. Imaging studies were obtained in 64 patients at least 12 months postsurgery (mean 19 months). Fusion was assessed by dynamic radiographs in all cases and with a CT scan in 80% of the cases. Four of the 64 patients did not achieve fusion (6.25% overall). All patients in whom fusion failed had undergone C1 lateral mass fixation and C2 pars (1/29, 3.4%) or laminar (3/9, 33.3%) fixation.

CONCLUSIONS Instrumented posterior cervical fusions may be performed in elderly patients with acceptable morbidity and mortality. The fusion rate is excellent except when bilateral C2 translaminar screws are used for axis fixation.

Spinal Robotics: Current Applications and Future Perspectives

Robotics Spinal Surgery

Neurosurgery 72:A12–A18, 2013

Even though robotic technology holds great potential for performing spinal surgery and advancing neurosurgical techniques, it is of utmost importance to establish its practicality and to demonstrate better clinical outcomes compared with traditional techniques, especially in the current cost-effective era. Several systems have proved to be safe and reliable in the execution of tasks on a routine basis, are commercially available, and are used for specific indications in spine surgery. However, workflow, usability, interdisciplinary setups, efficacy, and cost-effectiveness have to be proven prospectively.

This article includes a short description of robotic structures and workflow, followed by preliminary results of a randomized prospective study comparing conventional free-hand techniques with routine spine navigation and robotic-assisted procedures. Additionally, we present cases performed with a spinal robotic device, assessing not only the accuracy of the robotic-assisted procedure but also other factors (eg, minimal invasiveness, radiation dosage, and learning curves).

Currently, the use of robotics in spinal surgery greatly enhances the application of minimally invasive procedures by increasing accuracy and reducing radiation exposure for patients and surgeons compared with standard procedures. Second-generation hardware and software upgrades of existing devices will enhance workflow and intraoperative setup. As more studies are published in this field, robot-assisted therapies will gain wider acceptance in the near future.

Accuracy of Image-Guided Pedicle Screw Placement Using Intraoperative Computed Tomography- Based Navigation With Automated Referencing. Part II: Thoracolumbar Spine

Neurosurgery 69:1307–1316, 2011 DOI: 10.1227/NEU.0b013e31822ba190

Image-guided spinal instrumentation may reduce complications in spinal instrumentation.

OBJECTIVE: To assess accuracy, time efficiency, and staff radiation exposure during thoracolumbar screw instrumentation guided by intraoperative computed tomography (iCT)-based neuronavigation (iCT-N).

METHODS: In 55 patients treated for idiopathic and degenerative deformities, 826 screws were inserted in the thoracic (T2–T12; n = 243) and lumbosacral (L1–S1; n = 545) spine, as well as ilium (n = 38) guided by iCT-N. Up to 17 segments were instrumented following a single automated registration sequence with the dynamic reference arc (DRA) uniformly attached to L5. Accuracy of iCT-N was assessed by calculating angular deviations between individual navigated tool trajectories and final implant positions. Final screw positions were also graded according to established classification systems. Clinical and radiological outcome was assessed at 12 to 14 months.

RESULTS: Additional intraoperative fluoroscopy was unnecessary, eliminating staff radiation exposure. Unisegmental K-wire insertion required 4.6 6 2.9 minutes. Of the thoracic pedicle screws 98.4% were assigned grades I to III according to the Heary classification, with 1.6% grade IV placement. In the lumbar spine, 94.4% of screws were completely contained (Gertzbein classification grade 0), 4.6% displayed minor pedicle breaches ,2 mm(grade 1), and 1% of lumbar screws deviated by.2 to,4 mm (grade 2). The accuracy of iCT-N progressively deteriorates with increasing distance from the DRA, but allows safe instrumentation of up to 12 segments.

CONCLUSION: iCT-N using automated referencing allows for safe, highly accurate multilevel instrumentation of the entire thoracolumbosacral spine and ilium, rendering additional intraoperative imaging dispensable. In addition, automated registration is time-efficient and significantly reduces the need for re-registration in multilevel surgery.

Accuracy of Image-Guided Pedicle Screw Placement Using Intraoperative Computed Tomography- Based Navigation With Automated Referencing, Part I: Cervicothoracic Spine

Neurosurgery 69:782–795, 2011 DOI: 10.1227/NEU.0b013e318222ae16

Image-guided spinal instrumentation reduces the incidence of implant misplacement.

OBJECTIVE: To assess the accuracy of intraoperative computed tomography (iCT)-based neuronavigation (iCT-N).

METHODS: In 35 patients (age range, 18-87 years), a total of 248 pedicle screws were placed in the cervical (C1-C7) and upper and midthoracic (T1-T8) spine. An automated iCT registration sequence was used for multisegmental instrumentation, with the reference frame fixed to either a Mayfield head clamp and/or the most distal spinous process within the instrumentation. Pediculation was performed with navigated drill guides or Jamshidi cannulas. The angular deviation between navigated tool trajectory and final implant positions (evaluated on postinstrumentation iCT or postoperative CT scans) was calculated to assess the accuracy of iCT-N. Final screw positions were also graded according to established classification systems. Mean follow-up was 16.7 months.

RESULTS: Clinically significant screw misplacement or iCT-N failure mandating conversion to conventional technique did not occur. A total of 71.4% of patients self-rated their outcome as excellent or good at 12 months; 99.3% of cervical screws were compliant with Neo classification grades 0 and 1 (grade 2, 0.7%), and neurovascular injury did not occur. In addition, 97.8% of thoracic pedicle screws were assigned grades I to III of the Heary classification, with 2.2% grade IV placement. Accuracy of iCT-N progressively deteriorated with increasing distance from the spinal reference clamp but allowed safe instrumentation of up to 10 segments.

CONCLUSION: Image-guided spinal instrumentation using iCT-N with automated referencing allows safe, highly accurate multilevel instrumentation of the cervical and upper and midthoracic spine. In addition, iCT-N significantly reduces the need for reregistration in multilevel surgery.

Accuracy of pedicle screw placement in the lumbosacral spine using conventional technique: CT postoperative assessment in 102 consecutive patients

Journal of Neurosurgery: Spine. March 2010.DOI: 10.3171/2009.9.SPINE09261

The goal of this study was to determine the incidence of screw misplacement and complications in a group of 102 patients who underwent transpedicle screw fixation in the lumbosacral spine with conventional open technique and intraoperative fluoroscopy. The results are compared with published data.

Methods: Cases involving 102 consecutive patients (424 inserted screws) were reviewed. Surgery was performed in all cases by the same surgeon’s team, using the same implant, and all results were assessed by means of a specific CT protocol. The screw position was assessed by the authors and an independent observer. Screw position was classified as correct when the screw was completely surrounded by the pedicle cortex, as “cortical encroachment” (questionable violation) if the pedicle cortex could not be visualized, and as “frank penetration” when the screw was outside the pedicular boundaries. Frank penetration was further subdivided as minor (when the edge of the screw thread was up to 2.0 mm outside the pedicle cortex), moderate (2.1–4 mm), and severe (> 4 mm). The incidence of intra- and postoperative complications not related to screw position as well as hardware failures were also registered, with a minimum follow-up duration of 8 months.

Results: The rate of frank pedicle screw misplacement was 5%. The rate of minimal or questionable pedicle wall violation was 2.8%. Among the frank misplacements, 6 were classified as minor, 12 as moderate, and 3 as severe penetration. Two patients (2%) had radicular pain and neurological deficits (inferomedial and inferolateral minor misplacement at L-4 and L-5, respectively), and 5 patients (4.9%) complained only of radicular pain. At the follow-up examination all patients had completely recovered their neurological function and radicular pain was resolved in all cases. The complications not related to screw malposition were 2 pedicle fractures (2% of patients), 1 nerve root injury (1%), and 1 dural laceration (1%). Five patients (4.8%) had postoperative anemia and required transfusions. Superficial or deep wound infection was noted in 3 patients (2.9%). Late hardware failure occurred in 2 patients (2%). One patient developed adjacent segmental instability and required additional surgery to extend the fusion.

Conclusions: Our rates of screw misplacement and complications compare favorably with the lowest rates of the series in which conventional technique was used and are close to the rates reported for image-guided methods. The risk of malpositioning may be reduced with careful preoperative surgical planning, accurate knowledge of the spinal anatomy, surgical experience, and correct indication for conventional surgery. The conventional technique still remains a practical, safe, and effective surgical method for lumbosacral fixation.