Clinical accuracy and initial experience with augmented reality–assisted pedicle screw placement

J Neurosurg Spine 36:351–357, 2022

Augmented reality (AR) is a novel technology which, when applied to spine surgery, offers the potential for efficient, safe, and accurate placement of spinal instrumentation. The authors report the accuracy of the first 205 pedicle screws consecutively placed at their institution by using AR assistance with a unique head-mounted display (HMD) navigation system.

METHODS A retrospective review was performed of the first 28 consecutive patients who underwent AR-assisted pedicle screw placement in the thoracic, lumbar, and/or sacral spine at the authors’ institution. Clinical accuracy for each pedicle screw was graded using the Gertzbein-Robbins scale by an independent neuroradiologist working in a blinded fashion.

RESULTS Twenty-eight consecutive patients underwent thoracic, lumbar, or sacral pedicle screw placement with AR assistance. The median age at the time of surgery was 62.5 (IQR 13.8) years and the median body mass index was 31 (IQR 8.6) kg/m2. Indications for surgery included degenerative disease (n = 12, 43%); deformity correction (n = 12, 43%); tumor (n = 3, 11%); and trauma (n = 1, 4%). The majority of patients (n = 26, 93%) presented with low-back pain, 19 (68%) patients presented with radicular leg pain, and 10 (36%) patients had documented lower extremity weakness. A total of 205 screws were consecutively placed, with 112 (55%) placed in the lumbar spine, 67 (33%) in the thoracic spine, and 26 (13%) at S1. Screw placement accuracy was 98.5% for thoracic screws, 97.8% for lumbar/S1 screws, and 98.0% overall.

CONCLUSIONS AR depicted through a unique HMD is a novel and clinically accurate technology for the navigated insertion of pedicle screws. The authors describe the first 205 AR-assisted thoracic, lumbar, and sacral pedicle screws consecutively placed at their institution with an accuracy of 98.0% as determined by a Gertzbein-Robbins grade of A or B.

Augmented reality–mediated stereotactic navigation for execution of en bloc lumbar spondylectomy osteotomies

J Neurosurg Spine 34:700–705, 2021

En bloc spinal tumor resections are technically demanding procedures with high morbidity because of the conventionally large exposure area and aggressive resection goals. Stereotactic surgical navigation presents an opportunity to perform the smallest possible resection plan while still achieving an en bloc resection.

Augmented reality (AR)–mediated spine surgery (ARMSS) via a mounted display with an integrated tracking camera is a novel FDA-approved technology for intraoperative “heads up” neuronavigation, with the proposed advantages of increased precision, workflow efficiency, and cost-effectiveness. As surgical experience and capability with this technology grow, the potential for more technically demanding surgical applications arises.

Here, the authors describe the use of ARMSS for guidance in a unique osteotomy execution to achieve an en bloc wide marginal resection of an L1 chordoma through a posterior-only approach while avoiding a tumor capsule breach. A technique is described to simultaneously visualize the navigational guidance provided by the contralateral surgeon’s tracked pointer and the progress of the BoneScalpel aligned in parallel with the tracked instrument, providing maximum precision and safety. The procedure was completed by reconstruction performed with a quad-rod and cabled fibular strut allograft construct, and the patient did well postoperatively.

Finally, the authors review the technical aspects of the approach, as well as the applications and limitations of this new technology.

Spine Navigation Based on 3-Dimensional Robotic Fluoroscopy for Accurate Percutaneous Pedicle Screw Placement

World Neurosurg. (2017) 108:76-83.

Minimally invasive spine surgery is associated with obstructed visibility of anatomic landmarks and increased radiation exposure, leading to higher incidence of pedicle screw mispositioning. To address these drawbacks, intraoperative 3-dimensional fluoroscopy (io3DF) and navigation are being increasingly used. We aimed to present our dedicated multifunctional hybrid operating room (HyOR) setup and evaluate the accuracy and safety of io3DF image-guided spinal navigation in transforaminal lumbar interbody fusion with percutaneous pedicle screw (PPS) placement.

METHODS: The HyOR includes a fixed 3D multiaxis robotic fluoroscopy arm that moves automatically to the preprogrammed position when needed. An initial io3DF assessment is performed to collect intraoperative images, which are automatically transferred into the navigation system. These data are used to calibrate the PPSs and insert them under computer-assisted navigation. A second io3DF is performed for verifying PPS position.

RESULTS: Between January 2014 and December 2016, 66 consecutive patients (age, 58.6  14.1 years) were treated for refractory lumbar degenerative pain. Seventy-three spinal levels were treated, and 276 screws were placed, with 4.2 ± 0.76 screws per patient. There was no measurable radiation to the HyOR staff, whereas the mean radiation dose per patient was 378.3  uGym2. The overall accuracy rate of PPS placement was 99.6%. There were no significant procedure-related complications.

CONCLUSIONS: Spine navigation based on io3DF images enabled us to avoid radiation exposure to the operating room team while delivering minimal but sufficient radiation doses to our patients. This approach achieved an accuracy rate of 99.6% for PPS placement in the safe zone, without significant complications.