Intraoperative, full-rotation, three-dimensional image (O-arm)–based navigation system for cervical pedicle screw insertion

J Neurosurg Spine 15:472–478, 2011. DOI: 10.3171/2011.6.SPINE10809

The aim of this study was to retrospectively evaluate the reliability and accuracy of cervical pedicle screw (CPS) placement using an intraoperative, full-rotation, 3D image (O-arm)–based navigation system and to assess the advantages and disadvantages of the system.

Methods. The study involved 21 consecutive patients undergoing posterior stabilization surgery of the cervical spine between April and December 2009. The patients, in whom 108 CPSs had been inserted, underwent screw placement based on intraoperative 3D imaging and navigation using the O-arm system. Cervical pedicle screw positions were classified into 4 grades, according to pedicle-wall perforations, by using postoperative CT.

Results. Of the 108 CPSs, 96 (88.9%) were classified as Grade 0 (no perforation), 9 (8.3%) as Grade 1 (perforations < 2 mm, CPS exposed, and < 50% of screw diameter outside the pedicle), and 3 (2.8%) as Grade 2 (perforations between ≥ 2 and < 4 mm, CPS breached the pedicle wall, and > 50% of screw diameter outside the pedicle). No screw was classified as Grade 3 (perforation > 4 mm, complete perforation). No neurovascular complications occurred because of CPS placement.

Conclusions. The O-arm offers high-resolution 2D or 3D images, facilitates accurate and safe CPS insertion with high-quality navigation, and provides other substantial benefits for cervical spinal instrumentation. Even with current optimized technology, however, CPS perforation cannot be completely prevented, with 8.3% instances of minor violations, which do not cause significant complications, and 2.8% instances of major pedicle violations, which may cause catastrophic complications. Therefore, a combination of intraoperative 3D image–based navigation with other techniques may result in more accurate CPS placement.

Anterior single screw fixation of odontoid fracture with intraoperative iso-C 3-dimensional imaging

Eur Spine J (2011) 20:1899–1907. DOI 10.1007/s00586-011-1860-0
The purpose of this study was to assess the value of isocentric C-arm three-dimensional (Iso-C 3D) fluoroscopy for the insertion of an anterior odontoid screw. The results of the Iso-C 3D group were compared with that of an historic control group using conventional fluoroscopy.
Methods  Twenty-nine patients diagnosed with type II or rostral-type III odontoid fractures were treated with a single anterior screw fixation in this study. The Iso-C 3D group included 13 patients and the other 16 patients were in the historic control group. All operations were performed by a single surgeon using standard procedure and manner. The clinical and radiographic results were recorded and compared between the two groups.
Results  The fluoroscopy time in the Iso-C 3D group was 42.9 s as compared to 68.1 s in the control group (P\ 0.01). The mean operative time was 91.5 min in the Iso-C 3D group when compared with 81.6 min in the control group (P =  0.20). The rate of bony fusion was 96.6% (28/29), the failure rate of reduction or fixation was 13.8% (7.7% in Iso-C 3D group; 18.8% in control group). The Smiley–Webster scale showed that 90% of patients achieved good or better outcomes
Conclusions  In conclusion, this technique can be safely extended to the treatment of technically difficult to treat spinal injuries and at the same time reduce total radiation exposure time both for the patient and the surgeon.