Robot-Assisted Stereoelectroencephalography

Robot-Assisted Stereoelectroencephalography

Neurosurgery 78:169–180, 2016

Robot-assisted stereoelectroencephalography (SEEG) may represent a simplified, precise, and safe alternative to the more traditional SEEG techniques.

OBJECTIVE: To report our clinical experience with robotic SEEG implantation and to define its utility in the management of patients with medically refractory epilepsy.

METHODS: The prospective observational analyses included all patients with medically refractory focal epilepsy who underwent robot-assisted stereotactic placement of depth electrodes for extraoperative brain monitoring between November 2009 and May 2013. Technical nuances of the robotic implantation technique are presented, as well as an analysis of demographics, time of planning and procedure, seizure outcome, in vivo accuracy, and procedure-related complications.

RESULTS: One hundred patients underwent 101 robot-assisted SEEG procedures. Their mean age was 33.2 years. In total, 1245 depth electrodes were implanted. On average, 12.5 electrodes were implanted per patient. The time of implantation planning was 30 minutes on average (range, 15-60 minutes). The average operative time was 130 minutes (range, 45-160 minutes). In vivo accuracy (calculated in 500 trajectories) demonstrated a median entry point error of 1.2 mm (interquartile range, 0.78-1.83 mm) and a median target point error of 1.7 mm (interquartile range, 1.20-2.30 mm). Of the group of patients who underwent resective surgery (68 patients), 45 (66.2%) gained seizure freedom status. Mean follow-up was 18 months. The total complication rate was 4%.

CONCLUSION: The robotic SEEG technique and method were demonstrated to be safe, accurate, and efficient in anatomically defining the epileptogenic zone and subsequently promoting sustained seizure freedom status in patients with difficult-to- localize seizures.

Stereoelectroencephalography: Surgical Methodology, Safety, and Stereotactic Application Accuracy in 500 Procedures

Stereoelectroencephalography___Surgical

Neurosurgery 72:353–366, 2013

Stereoelectroencephalography (SEEG) methodology, originally developed by Talairach and Bancaud, is progressively gaining popularity for the presurgical invasive evaluation of drug-resistant epilepsies.

OBJECTIVE: To describe recent SEEG methodological implementations carried out in our center, to evaluate safety, and to analyze in vivo application accuracy in a consecutive series of 500 procedures with a total of 6496 implanted electrodes.

METHODS: Four hundred nineteen procedures were performed with the traditional 2- step surgical workflow, which was modified for the subsequent 81 procedures. The new workflow entailed acquisition of brain 3-dimensional angiography and magnetic resonance imaging in frameless and markerless conditions, advanced multimodal planning, and robot-assisted implantation. Quantitative analysis for in vivo entry point and target point localization error was performed on a sub–data set of 118 procedures (1567 electrodes).

RESULTS: The methodology allowed successful implantation in all cases. Major complication rate was 12 of 500 (2.4%), including 1 death for indirect morbidity. Median entry point localization error was 1.43 mm (interquartile range, 0.91-2.21 mm) with the traditional workflow and 0.78 mm (interquartile range, 0.49-1.08 mm) with the new one (P , 2.2 · 10216). Median target point localization errors were 2.69 mm (interquartile range, 1.89-3.67 mm) and 1.77 mm (interquartile range, 1.25-2.51 mm; P, 2.2 · 10216), respectively.

CONCLUSION: SEEG is a safe and accurate procedure for the invasive assessment of the epileptogenic zone. Traditional Talairach methodology, implemented by multimodal planning and robot-assisted surgery, allows direct electrical recording from superficial and deep-seated brain structures, providing essential information in the most complex cases of drug-resistant epilepsy.

Stereoelectroencephalography: Surgical Methodology, Safety, and Stereotactic Application Accuracy in 500 Procedures

Stereoelectroencephalography- Surgical Methodology, Safety, and Stereotactic Application Accuracy in 500 Procedures

Neurosurgery 72:353–366, 2013

Stereoelectroencephalography (SEEG) methodology, originally developed by Talairach and Bancaud, is progressively gaining popularity for the presurgical invasive evaluation of drug-resistant epilepsies.

OBJECTIVE: To describe recent SEEG methodological implementations carried out in our center, to evaluate safety, and to analyze in vivo application accuracy in a consecutive series of 500 procedures with a total of 6496 implanted electrodes.

METHODS: Four hundred nineteen procedures were performed with the traditional 2- step surgical workflow, which was modified for the subsequent 81 procedures. The new workflow entailed acquisition of brain 3-dimensional angiography and magnetic resonance imaging in frameless and markerless conditions, advanced multimodal planning, and robot-assisted implantation. Quantitative analysis for in vivo entry point and target point localization error was performed on a sub–data set of 118 procedures (1567 electrodes).

RESULTS: The methodology allowed successful implantation in all cases. Major complication rate was 12 of 500 (2.4%), including 1 death for indirect morbidity. Median entry point localization error was 1.43 mm (interquartile range, 0.91-2.21 mm) with the traditional workflow and 0.78 mm (interquartile range, 0.49-1.08 mm) with the new one (P < 2.2 · 10*16). Median target point localization errors were 2.69 mm (interquartile range, 1.89-3.67 mm) and 1.77 mm (interquartile range, 1.25-2.51 mm; P< 2.2 · 10*16), respectively.

CONCLUSION: SEEG is a safe and accurate procedure for the invasive assessment of the epileptogenic zone. Traditional Talairach methodology, implemented by multimodal planning and robot-assisted surgery, allows direct electrical recording from superficial and deep-seated brain structures, providing essential information in the most complex cases of drug-resistant epilepsy.