Novel Delivery System Significantly Improves Technical Success and Safety of the Endovascular Clip System (eCLIPs) While Maintaining Efficacy for the Treatment of Wide-Necked Bifurcation Aneurysms

Neurosurgery 97:963–970, 2025

A prospective registry study found that the new eCLIPs electrolytic bifurcation system (eB) for wide-necked bifurcation aneurysms significantly improved technical success and safety over the prior generation, while maintaining high efficacy, with 96% satisfactory occlusion and low rates of neurological complications and retreatment.

• The eCLIPs device is an endovascular implant designed to treat wide-necked bifurcation aneurysms (WNBAs).

• This study compares two generations: the first-generation eCLIPs bifurcation remodeling system (eBRS) and the newer eCLIPs electrolytic bifurcation system (eB).

• A prospective registry included 280 patients; the eB system showed higher implantation success (92% vs 81% for eBRS, P = .007).

• Both systems achieved high rates of satisfactory aneurysm occlusion (mRROC 1+2: 96%).

• The eB system had lower all-cause (1.1%) and neurological (0.5%) death rates compared to eBRS (7% and 5%, respectively).

• Stroke and repeat procedure rates were low for both devices, with eB performing better (stroke: ~1.7%, repeat procedure: 2.2% vs 8.9% for eBRS).

• eCLIPs compares favorably in safety and efficacy to other devices like WEB and stent-assisted coiling.

• The eB system offers improved procedural success and safety, making eCLIPs a strong option for WNBAs; further studies are recommended.