The Spine Journal 24 (2024) 1671−1677
Accurate pedicle screw placement is a challenge with reported misplacement rates of 10% and higher. A handheld navigation device (HND) may provide accuracy equal to CT-based navigation (CT-Nav) but without the cost and complexity.
OBJECTIVE: To study the accuracy of a handheld navigation device for pedicle screw placement. STUDY DESIGN: This prospective cross-sectional study with consistently applied reference standard enrolled 20 patients undergoing 92 pedicle screw placements.
PATIENTS: Patients who underwent pedicle screw placement between May 2022 and September 2022.
OUTCOME MEASURES: Pedicle screw placement accuracy per Gertzbein−Robbins.
METHODS: Once the screw pilot hole was established, the proposed trajectory of the HND was compared with that proposed by CT-Nav. Postoperatively, screw accuracy was graded according to Gertzbein−Robbins by a blinded radiologist based on CT scans. Accuracy was compared between the two systems and published control for fluoroscopy assisted and CT-Nav placement using Bayesian posterior distribution.
RESULTS: The trajectory proposed by the HND and CT-Nav were in agreement in 98.9% (95% Exact CI; 94.09%−99.97%). The HND accuracy was 98.9% with 91 screws rated “A” and 1 rated “C”. Noninferiority to fluoroscopic placement was achieved because the one-sided normalapproximation 95% CI Lower Bound (LB) of 95.3% is greater than the Performance Goal (PG) of 83.4%. Posthoc analysis demonstrated that the probability of superiority of the HND relative to the historical accuracy rate of 91.5% for fluoroscopy assisted procedures is >0.999 and that the HND’s accuracy rate is within 4.5% of CT-Nav of 95.5% is >0.999. No adverse events or intra-operative complications associated with HND were observed. There was 1 (1.1%) intra-operative repositioning and no reoperations for any reason.
CONCLUSIONS: The accuracy rate of the HND was 98.9%, and the proposed trajectory matched with CT-Nav in 98.9% of the time. This is superior to the historical published accuracy rate for fluoroscopy-assisted procedures and equivalent to the historical published accuracy rate for CT-Nav.